Amptek COOL-X Operating Manual Download Page 2

AMPTEK INC. 

                                                       COOL-X  Operating Manual 

Page 2 

Radiation Precautions 

 

The COOL-X is intended to generate x-rays up to 35 keV during normal heating 
and cooling phases.  However, ambient or room temperature fluctuations will also 
result in producing a small amount of x-rays. 
 

WHEN NOT IN USE OR IN STORAGE, ALWAYS KEEP THE 
PROTECTIVE METAL COVER ATTACHED TO THE FRONT OF THE 
COOL-X. 

 
The COOL-X normally emits x-rays in the forward output direction.  However, 
radiation in other directions is possible and should be addressed with shielding 
and/or monitoring in the final application. 
 

Caution   

This device produces X-Rays when energized. 

To be operated only by qualified personnel.  

 

Radiation levels external to the x-ray generator housing with the cover on does 
not exceed 2.5 mrem/h measured 5 cm from the surface of the housing in 
accordance with Requirements 5.2.2.1.1 and 5.2.2.2.2 of the NBS Handbook for 
Radiation Safety for X-Ray Diffraction and Fluorescence Analysis Equipment 
(for more information, see http://www.fda.gov/cdrh/radhlth/pdf/xdfgehb.pdf). 

 
 

Beryllium Window Caution    

 

The front of the COOL-X is made of a 10 mils (250 µm) Beryllium window. 
 
DO NOT REMOVE the metal cover from the COOL-X until ready to use.  The 
Be window is extremely brittle and can shatter easily.  Do not have any object 
come into contact with the window.  Do not touch the window because oil from 
the fingers will cause it to oxidize.  The window cannot be repaired.  If the 
window is broken the COOL-X element must be replaced.   
 

Beryllium windows damaged by improper handling will not be covered by 
the warranty. 

 

 

Caution 

The COOL-X is only one component of an X-ray instrument.  It is the 
responsibility of the user, the OEM customer, or experimenter to provide a fail-
safe metal enclosure to prevent escaping radiation while using this product.  The 
final product (turn-key system) must comply with local government regulations to 
protect personnel from exposure to radiation.  Amptek Inc. bears no responsibility 
for the incorrect use of this product. 

Summary of Contents for COOL-X

Page 1: ...COOL X X Ray Generator with Pyroelectric Crystal Operating Manual AMPTEK INC 14 DeAngelo Drive Bedford MA 01730 2204 USA 1 781 275 2242 Fax 1 781 275 3470 Amptek sales ametek com www amptek com ...

Page 2: ...tion Safety for X Ray Diffraction and Fluorescence Analysis Equipment for more information see http www fda gov cdrh radhlth pdf xdfgehb pdf Beryllium Window Caution The front of the COOL X is made of a 10 mils 250 µm Beryllium window DO NOT REMOVE the metal cover from the COOL X until ready to use The Be window is extremely brittle and can shatter easily Do not have any object come into contact w...

Page 3: ... defects or other reproductive harm if exposed to them through improper use storage or disposal of the product Prop 65 Chemical Type of Toxicity CAS No Product part containing the chemical Beryllium Cancer Detector window Please consult this owner s manual for proper use storage care and disposal of the product For more information go to www p65warnings ca gov ...

Page 4: ...nerator is completely enclosed inside the small TO 8 package at the tip of the extender TURN OFF POWER TO THE COOL X WHEN NOT IN USE 2 The Controller is enclosed inside the metal box The function of the Controller is to supply power to the X ray generator monitor its temperature and automatically change the heating and cooling phases of the cycle Figure 1 Please see Appendix A if the COOL X emitte...

Page 5: ... is with the AC adapter If battery operation is required remove the 4 small screws from the top of the box and insert the provided 9 Volt battery to the terminals 4 The COOL X should be mounted on a firm surface using the 4 mounting tabs of the box 5 Connect the AC adapter to the back of the box and remove the front metal cover Power On Off AC Adapter Jack input Flashing LED Figure 2 Figure 3 ...

Page 6: ...unit is toggled ON OFF every time the power button is pressed When the power is OFF the LED is also OFF 10 At the start of a cycle and in between the heating cooling phases a BEEP sound will be heard from a small internal speaker to indicate the transition of phases If the sound interferes with the x ray detector due to microphonic pick up the speaker can be disconnected To disconnect the speaker ...

Page 7: ...tory This warranty does not apply in the event of misuse or abuse of the instrument or unauthorized alterations or repair AMPTEK INC shall not be liable for any consequential damages including without limitation damages resulting from the loss of use due to failure of this instrument The factory MUST be notified prior to return shipment All products returned under the warranty must be shipped prep...

Page 8: ... pattern for the TO 8 for attaching it to the heat sink is provided below The Control electronics should also be mounted on a metal plate and kept at approximately the same temperature as the TO 8 This is important because the temperature sensor in the Control circuit refers to a matched temperature sensor inside the TO 8 emitter head in order to maintain the proper temperature differentials and t...

Reviews: