8
Continuity Test (See Figure 5)
1. Set the Function Switch to
R
.
2. Connect the test leads: Red to
Temp
V
Ω
Hz
, Black to
COM
.
3. Turn off power to the circuit being measured.
4. Discharge any capacitors that may influence the reading.
5. Connect the test leads across the resistance.
6. Listen for the tone that indicates continuity (< 40
Ω
).
When measuring continuity
the meter emits a continuous tone when the resistance value
falls below 40
Ω
.
Resistance Measurement (See Figure 6)
1. Set the Function Switch to
Ω
.
2. If
RANGE
is displayed, press the
RANGE
button to enable auto ranging.
3. Connect the test leads. Red to
Temp
V
Ω
Hz
, Black to
COM
.
4. Turn off power to the circuit being measured. Never measure resistance across a
voltage source or on a powered circuit.
5. Discharge any capacitors that may influence the reading.
6. Connect the test leads across the resistance.
7. Read the display. If OL appears on the highest range, the resistance is too large to
be measured.
Note:
When measuring very low resistances, use Relative Measurement to eliminate the
test lead resistance (see Button Functions later in this manual).
Summary of Contents for HD160C
Page 2: ......
Page 5: ...HD160C Heavy Duty True rms Digital Multimeter User Manual English ...
Page 27: ......
Page 28: ......
Page 29: ...HD160C Heavy Duty True rms Digital Multimeter Mode d emploi Francais ...
Page 51: ......
Page 52: ......
Page 53: ...HD160C Heavy Duty True rms Digital Multimeter Bedienungshandbuch Deutsch ...
Page 75: ......
Page 76: ......
Page 77: ...HD160C Heavy Duty True rms Digital Multimeter Manuale d uso Italiano Italiano Italiano ...
Page 99: ......
Page 100: ......
Page 101: ...Español HD160C Heavy Duty True rms Digital Multimeter Manual de uso ...
Page 123: ......
Page 124: ......
Page 125: ...Svenska HD160C Heavy Duty True rms Digital Multimeter Anvandärhandbok ...
Page 147: ......
Page 148: ......