Amprobe AT-8000 Series Quick Start Manual Download Page 5

TIP:

  In Energized mode, align the groove on the Tip 

Sensor with the wire direction for best results; the 

signal may not be detected without this alignment.  
De-energized mode uses a different antenna in the 

Tip Sensor than Energized mode. Specific alignment 

of the Tip Sensor groove to the wire is not required. 

De-energized wire tracing results are based only on 

how close the Tip Sensor is to the wire.

 ©2020 Amprobe

®

 For detailed specifications and ordering go to amprobe.com  6012404 A

3

Advanced Wire Tracer - Quick Guide

Tracing Energized and De-energized Wires - Tip Sensor

Receiver: 

Energized and De-energized 

Tip Sensor Modes

Use this mode for pinpointing a wire in a bundle or 
tracing in corners and confined spaces such as junction 
boxes and inside enclosures.

1.  Turn on

 the Receiver and select either Energized 

or De-energized TIP SENSOR mode using the 

directional arrows.

2.  Hold 

the Receiver with the Tip Sensor facing the 

target area.

3.  Scan

 the target area with the Tip Sensor to find the 

highest signal level, then begin tracing the detected 

wire. Increase or decrease sensitivity of the Receiver 

by pr or - on the keypad as necessary.

4.  Press ENTER

 when complete to return to the 

home screen.

Receiver screen showing signal detected in 

Energized TIP SENSOR mode

30-600V AC/DC

40-400 HZ

TIP SENSOR 

ENERGIZED

75

SIGNAL

Aligning the Tip Sensor with the wire

Tip groove

Align

AT-8000-T

Set-up: 

Test Leads

Set-up: 

Transmitter

1.  Plug

 the green and red test leads into the Transmitter.

2.  Connect

 the green wire to a separate ground. 

3.  Connect

 the red test lead to the wire being traced. 

For receptacles, make sure to connect the test lead to 

the line (hot) wire. For Energized systems the signal 

will ONLY be transmitted between the load-side to 

which the Transmitter is connected and the source 

of power.

1.  Turn on

 the Transmitter.

2.  Verify

 that the test leads are properly connected; 

the red LED voltage status light should be on 

for circuits with voltage above 30 V AC/DC, and 

it should be off for De-energized circuits below 

30 V AC/DC.

3.  Select HIGH signal mode

 by pressing HI.

2

1

Proper connection 

with separate 

ground

AT-8000-T

TRANSMITTER

HI

LO

Blinking

Transmitter 

indicator showing 

signal in  

HIGH mode

Summary of Contents for AT-8000 Series

Page 1: ...Fluke and safety certified by 3rd party labs Intuitive Transmitter automatically senses whether the system is energized or de energized Most accurate wire tracing in its class with 10 sensitivity mode...

Page 2: ...AT 8000 Advanced Wire Tracer...

Page 3: ...GFCI protected circuit wire tracing Find breaks openings and shorts Trace wires in metal conduit junction box method Trace non metallic pipes and conduits Trace shielded wires Trace underground wires...

Page 4: ...onnect the test lead to the line hot wire For Energized systems the signal will ONLY be transmitted between the load side to which the Transmitter is connected and the source of power 1 Turn on the Tr...

Page 5: ...ignal level then begin tracing the detected wire Increase or decrease sensitivity of the Receiver by pressing or on the keypad as necessary 4 Press ENTER when complete to return to the home screen Rec...

Page 6: ...3 Connect the red test lead to the wire being traced For receptacles make sure to connect the test lead to the line hot wire For Energized systems the signal will ONLY be transmitted between the load...

Page 7: ...Receiver and select either Energized BREAKERS mode or De Energized BREAKERS mode using the directional arrows 2 Align the groove on the Tip Sensor with the breaker lengthwise 3 Scan each breaker by t...

Page 8: ...is necessary 1 Turn on the Receiver and press the NCV button 2 Hold the Receiver with the Tip Sensor facing the target area 3 Scan the target area with the Tip Sensor to find the highest signal level...

Page 9: ...place of the fault move the Transmitter to the other end of the wire and repeat tracing from the opposite end If the signal stops at the exact same location the fault has been located 1 Turn on the Tr...

Page 10: ...ed circuits below 30 V AC DC 3 Turn the Transmitter to Loop mode by pressing HIGH button for two seconds Verify that the Loop LED is ON 1 2 Note This method will be affected by signal cancellation eff...

Page 11: ...5cm radius 5 Detection Up to 4 FT 121 92cm 5 N A N A Display Size 3 5 in 89 mm LEDs N A Display Dimensions W x H 2 76 x 2 07 in 70 x 52 mm N A N A Display Resolution 320 x 240 N A N A Display Type Col...

Page 12: ...g IP 20 Drop Test 3 28 ft 1 m Dimensions Red black leads 3 28 ft 1 m Green lead 22 97 ft 7 m Alligator clips approx 3 74 x 1 77 x 0 94 in 95 x 45 x 24 mm Outlet adapters 2 83 x 0 71 x 0 71 in 72 x 18...

Page 13: ...2020 Amprobe For detailed specifications and ordering go to amprobe com 6012404 A 11 Advanced Wire Tracer Quick Guide...

Page 14: ...Traceur de fils avanc AT 8000...

Page 15: ...es circuits ouverts et des courts circuits Tracer des fils dans un conduit m tallique m thode avec bo tier de jonction Tracer des tuyaux et des conduits non m talliques Tracer des fils blind s Tracer...

Page 16: ...c ble d essai au fil de ligne aliment Sur les syst mes sous tension le signal sera UNIQUEMENT transmis entre le c t charge auquel le transmetteur est raccord et la source d alimentation 1 Allumez le...

Page 17: ...ur de pointe pour trouver le niveau de signal le plus lev puis commencez le tra age du fil d tect Augmentez ou diminuez la sensibilit du r cepteur en appuyant sur ou sur le clavier si n cessaire 4 App...

Page 18: ...e au fil trac Pour les prises lectriques assurez vous de raccorder le c ble d essai au fil de ligne aliment Sur les syst mes sous tension le signal sera UNIQUEMENT transmis entre le c t charge auquel...

Page 19: ...1 Allumez le r cepteur et s lectionnez le mode DISJONCTEURS sous tension ou DISJONCTEURS hors tension l aide des fl ches directionnelles 2 Alignez la rainure du capteur de pointe avec la longueur du...

Page 20: ...le bouton NCV 2 Tenez le r cepteur avec le capteur de pointe orient vers la zone cible 3 Balayez la zone cible avec le capteur de pointe pour trouver le niveau de signal le plus lev puis commencez le...

Page 21: ...signal s arr te 2 Pour v rifier la localisation du d faut d placez le transmetteur l autre extr mit du fil et r p tez le tra age depuis l extr mit oppos e Si le signal s arr te exactement au m me endr...

Page 22: ...tension avec une tension inf rieure 30 V CA CC 3 Mettez le transmetteur en mode Boucle en appuyant sur le bouton LEV pendant 2 secondes V rifiez que la LED Boucle est allum e 1 2 Remarque Cette m thod...

Page 23: ...i 152 4 cm 2 Capteur de pointe Sous tension Rep rage Environ 1 97 po 5 cm 1 D tection Jusqu 22 pi 670 56cm 1 Capteur de pointe Hors tension D tection Jusqu 14 pi 426 72cm 5 NCV 40 400 Hz Rep rage Envi...

Page 24: ...8 pi 1 m Dimensions Fils rouge noir 1 m Fil vert 7 m Pinces crocodile environ 95 x 45 x 24 mm Adaptateurs de sortie 72 x 18 x 18 mm Poids Environ 0 88 lb 0 4 kg Certifications Le kit c ble d essai et...

Page 25: ...2020 Amprobe Pour obtenir les sp cifications d taill es et commander rendez vous sur amprobe com 6012404 A 11 S rie de traceurs de fils avanc s Guide rapide...

Page 26: ...AT 8000 Rastreador de cable avanzado...

Page 27: ...de rupturas circuitos abiertos y cortocircuitos Rastreo de cables en conductos met licos m todo de caja de empalmes Rastreo de tuber as y conductos no met licos Rastreo de cables blindados Rastreo de...

Page 28: ...de l nea vivo Para los sistemas energizados la se al SOLO se transmitir nicamente entre el lado de carga al que est conectado el transmisor y la fuente de alimentaci n 1 Encienda el transmisor 2 Verif...

Page 29: ...alto y a continuaci n comience con el rastreo del cable detectado Aumente o reduzca la sensibilidad del receptor presionando o en el teclado seg n sea necesario 4 Presione ENTER INTRO cuando haya fina...

Page 30: ...n el caso de recept culos aseg rese de conectar el terminal de prueba al cable de l nea vivo Para los sistemas energizados la se al SOLO se transmitir nicamente entre el lado de carga al que est conec...

Page 31: ...LOCALIZAR INTERRUPTORES FUSIBLES INTERRUPTORES ENERGIZADOS INTERRUPTORES FUSIBLE ENCONTRADO EXPLORAR LOCALIZAR Paso 1 1 LECTURA 1 Encienda el receptor y seleccione el modo de DISYUNTORES energizados...

Page 32: ...eptor con el sensor de punta con orientaci n hacia el rea de destino 3 Realice la lectura de un rea deseada con el sensor de punta para buscar el nivel de se al m s alto y a continuaci n comience con...

Page 33: ...a que se detenga la se al 2 Verifique el lugar de la falla mueva el transmisor al otro extremo del cable y repita el rastreo del extremo opuesto Si la se al se detiene en la misma ubicaci n exacta ha...

Page 34: ...s desenergizados por debajo de 30 V de CA CC 3 Coloque el transmisor en el modo Bucle presionando brevemente HIGH ALTO durante dos segundos Verifique que el LED de bucle est encendido 1 2 Nota Este m...

Page 35: ...punta Desenergizado Detecci n Hasta 426 72 cm 14 pies 5 NCV 40 a 400 Hz Ubicaci n con precisi n Alrededor de 1 97 pulg 5 cm de radio 5 Detecci n Hasta 121 92 cm 4 pies 5 N D N D Tama o de pantalla 89...

Page 36: ...rojo negro 3 28 pies 1 m Terminal verde 22 97 pies 7 metros Pinzas de cocodrilo aprox 3 74 x 1 77 x 0 94 pulgadas 95 x 45 x 24 mm Adaptador de tomacorrientes 2 83 x 0 71 x 0 71 pulgadas 72 x 18 x 18...

Page 37: ...2020 Amprobe Para acceder a las especificaciones detalladas y para solicitar el producto dir jase a amprobe com 6012404 A 11 Serie de rastreadores de cable avanzados gu a r pida...

Page 38: ...Fluke Corp 6920 Seaway Blvd M S 143F Everett WA 98203 USA amprobe com 877 AMPROBE 267 7623 2019 2020 Fluke Corporation All rights reserved Printed in the U S A 6012404 A Specifications subject to cha...

Reviews: