Additional Features
Input Test Lead Warning
The meter emits a continuous tone when a test lead is placed in the
mA
or
10A
input jack and the Function/Range Switch is not set to a correct current position. (If
the meter is connected to a voltage source with leads connected for current, very
high current could result). All current ranges are protected by fast acting fuses.
True-rms Measurements
For ac measurements most DMMs average the ac input signal and display the
result as an estimated rms value. This average-responding method is accurate for
sinusoidal waveforms, but can be very inaccurate for distorted waveforms. To
ensure the most accurate measurements, always use a true-rms DMM when
measuring ac voltage or ac current on circuits for the following kinds of
applications:
x
Power Supplies - diodes
x
Controllers
x
Power Limiting - SCR or Triac
x
Starting - motors
x
Florescent Lighting - ballasts
x
Speed Control - motors
x
Pulsed Signals
x
Any non-sinusoidal ac waveform
MIN MAX AVG Measurements
The MIN MAX AVG function reads and updates the display to show the maximum
or minimum value measured after you press the
MIN MAX AVG
button.
Pressing the
MIN MAX AVG
button for less than 1 second will put the meter into a
mode of displaying the maximum, minimum, average, or actual readings. Each time
the button is pressed, the meter will cycle to the next display mode as shown in the
table below. Press the
MIN MAX AVG
button for more than 2 seconds to disable
this feature.
Button
Display
Value Displayed
< 1 second
REC MAX
Maximum value after feature activated
REC MIN
Minimum value after feature activated
< 1 second
REC AVG
Average value after feature is activated
< 1 second
REC
Actual reading, min max being recorded.
> 2 seconds
Exit
MIN MAX AVG
Normal measurement, actual reading
Peak Hold Measurements
Note: The PEAK function calibrates itself to meet the specifications.
Peak Hold records and stores the positive and negative peak values that occur while
measuring ac current or ac voltage. To enable the Peak Hold feature press the
PEAK
button for more than 2 seconds. The display will show
CAL
to indicate the
calibration cycle is in process. After the
CAL
indication clears, press the
PEAK
button again to display the maximum (P+) value for the ac voltage or ac current
being measured. The display will toggle between the P+ and P- readings each time
the
PEAK
button is pressed. Press the
PEAK
button for more than 1 second to exit
the PEAK function.
Summary of Contents for 37XR-A
Page 1: ......
Page 4: ......
Page 18: ...14 5 4 1 4 3 1 1 2 3 2 6 V V 2 5 37XR A 37XR A ...
Page 20: ...16 5 4 7 5 6 5 8 6 7 A 1 3 4 1 4 2 2 8 A 3 9 37XR A 37XR A ...
Page 21: ...17 5 7 6 6 4 5 7 4 1 6 7 3 1 2 2 3 8 37XR A 37XR A ...
Page 22: ...18 2 8 1 3 5 4 1 5 6 7 3 9 4 8 2 6 37XR A 37XR A ...
Page 23: ...19 11 1 5 6 4 3 1 10 5 4 2 6 3 2 7 37XR A 37XR A ...
Page 24: ...20 3 5 4 3 12 13 4 5 1 2 2 1 37XR A 37XR A ...
Page 25: ...21 3 4 5 1 2 14 2 4 2 15 6 37XR A ...
Page 26: ...22 ...
Page 40: ...14 ...
Page 54: ...14 ...
Page 68: ...14 ...
Page 82: ...14 ...
Page 83: ...C M Y CM MY CY CMY K ...