TAKING MEASUREMENTS
15
D-0115692-C (OM031)
–
Otowave 202 Operating Manual
TESTING LEFT EAR
Equalising Pressure
Cancel
TESTING LEFT EAR
Pressure Settling
Cancel
Press
◄ at any time to cancel the test and return to the ear selection menu.
TESTING LEFT EAR
Seal Obtained
Taking Tympanogram
Cancel
Once an adequate seal is detected the tympanogram measurement is made. This takes about 3 seconds. It is
important not to move the probe and to ask the patient to remain very still during the test.
When the tympanogram is complete the instrument will perform the reflex test(s), if selected. By default this
test is only performed if a peak is found in the tympanogram. This and other reflex test options may be changed
in the CONFIGURATION menu, see Sections 6 and 5.7.
Before starting the reflex test the ear canal pressure will be set to the value that gave the peak admittance
during the tympanogram test. The instrument will then step through the tone frequencies and levels set in the
CONFIGURATION menu searching for a reflex response. If selected, an ipsilateral reflex is tested first:
TESTING LEFT EAR
Seeking Ipsi Reflex
1000 Hz 80 dB
Cancel
The display changes to show the frequency and level being used, starting with the lowest frequency and level
selected.
This will be followed by a contralateral reflex test if this has been selected, with the display showing the
frequency and level being used: