TAKING MEASUREMENTS
14
D-0115692-C (OM031)
–
Otowave 202 Operating Manual
Ensure that the ear tip is pushed all the way down on the
probe tip and that there is no gab between probe tip and ear
tip.
The small holes through the Otowave probe tip must be kept
clear. If these become blocked a warning message will be
displayed. The probe tip must be cleaned or removed and
replaced.
5.3.
PERFORMING A TEST
Having selected the required test settings a typical tympanogram measurement and reflex tests are carried
out as follows.
From the MAIN MENU select NEW TEST:
MAIN MENU
NEW TEST
CONFIGURATION
VIEW THE LAST TEST
Select
Select the ear(s) required for test:
SELECT EAR
BOTH
LEFT
RIGHT
Back
Select
The message “Deleting last test” will be displayed momentarily followed by a message to insert the probe into
the ear to be tested:
TESTING LEFT EAR
INSERT PROBE
Cancel
Present the ear tip to the ear and obtain a seal. If a good seal has been detected the following sequence of
messages will be seen