270+ Operating Manual
5.2.
CLEANING THE 270+
•
Use caution while cleaning.
•
Before cleaning, remove the 270+ from mains power.
•
Do not allow any liquid to enter any part of the instrument or accessories.
•
Do not autoclave or sterilise the instrument or any accessories.
•
Do not use hard, sharp or pointed objects to clean any part of the instrument or an accessories.
•
If parts have been in contact with fluids do not allow them to dry before cleaning.
•
Follow local best practice and safety guidelines if available.
•
Clean the instrument by wiping the outer case with a lint free cloth lightly dampened with cleaning
solution. Recommended cleaning and disinfection solutions are warm water with mild, nonabrasive
cleaning solution (soap) and/or Clinical wipes (for example Clinell Universal).
•
If disinfection is required, use a disinfectant wipe rather than a spray product. Make sure that excess
liquid from the wipe does not seep into any sensitive areas such as connectors and seams where
plastic pieces connect such as under the rubber buttons on the 270+. Follow the instructions on the
disinfection product.
5.3.
CLEANING THE ACCESSORIES
5.3.1.
TRANSDUCER MAINTENANCE
Before use check the transducer cables and connectors for signs of wear and/or damage. If you find any,
please replace the item immediately by contacting Amplivox or your Amplivox distributor, requesting the
relevant part number.
Handle the audiometric headset, bone vibrator headset and other accessories with care. For parts that are in
direct contact with the patient it is recommended that replacement parts are used or the parts are subjected to
a standard disinfecting procedure between patients.
This includes physically cleaning and use of a recognised disinfectant. The specific manufacturer's instructions
should be followed for use of this disinfecting agent to provide an appropriated level of cleanliness.
During the cleaning process do not allow moisture to enter the earphone, insert masker, monitor or
microphone grills etc. For specific accessories refer to the sections below.
5.3.2.
EARPHONES
Clean the ear cushions (including those on the Audiocups, if used) with a recognised disinfectant, e.g. a
“Mediswab”.
5.3.3.
INSERT MASKER
Never insert or in any way use the insert masker without using a new, clean and fault-free test tip. This part is
for single use only - that is, each test tip is intended to be used once only for a single ear for a single patient.
Do not reuse test tips as this will pose the risk of ear-to-ear or patient-to-patient cross infection.
5.3.4.
INSERT EARPHONES
The disposable foam eartips supplied with the optional EarTone5A insert transducers are for single use only -
that is, each eartip is intended to be used once only for a single ear for a single patient. Do not reuse eartips
as this will pose the risk of ear-to-ear or patient-to-patient cross infection.
CAUTION