Amphenol Wilcoxon TN14 Troubleshooting Installations Download Page 6

Re-measuring with the sensor mounted at a different location will usually discriminate machine vibration from 

mounting resonance. However if a machine signal coincides with the resonance of the mounting configuration, 

a greatly increased signal can happen.  This can lead to sensor overload which will be manifested with a spec-

trum dominated by a ski slope.

Line frequency harmonics in spectrum

Harmonics of AC line power frequency usually indicate interference from motors, power lines and other emis-

sive equipment. First, ensure that the sensor shield is grounded (at one end only). If the shielding is good, 

check the cable routing. A good measurement practice is to run signal cables in cable trays separate from 

power line cable trays by at least 12 inches.  If a signal cable tray must come near a power cable tray, they 

should cross at right angles to reduce the possibility of magnetically coupled signals. Using shielded, twisted 

pair cable will also help minimize any magnetically coupled noise into the signal cable.  For example, if a power 

cable is 440 volts and the vibration signals from the sensor are at the millivolt level, any cross talk will severely 

corrupt the data.

Mounting resonance spectrum

Mounting resonance can give false indication of high frequency machinery faults such as gear mesh and bear-

ing problems. The problem is more likely to occur when using probe tips and magnets, and care must be exer-

cised to prevent the measured mounting resonance peaks from being falsely identified as machine produced 

peaks. Mounting the sensor on thin plates such as machine guards, can also lower the mounting resonance. 

Figure 8 shows the resonance of several common mounting techniques. Mounting resonance plots are shown 

for: (a) probe tip; (b) magnet; (c) Wilcoxon’s QuickLINK™; and (d) stud mounted configurations.

Figure 8a: Probe tip mounting resonance

Figure 8b: Magnet mounting resonance

Figure 8c: QuickLINK

TM

 mounting resonance

Figure 8d: Stud mounting resonance

Page 6

Summary of Contents for Wilcoxon TN14

Page 1: ...capacitor in the measurement equipment leaving the AC output signal Most vibration data collectors monitors and sensor power units contain an internal blocking capacitor for AC coupling If not includ...

Page 2: ...a true analog representation of the vibration the sensor is attempting to measure Measuring the BOV The constant current diode CCD limits the current supplied to the sensor It provides a constant cur...

Page 3: ...een integrated to velocity or displacement may also produce ski slopes for various reasons Cable routing faults can also be detected by analyzing the FFT Multiples of the line power frequency usually...

Page 4: ...dustrial sensors typically contain protection devices to prevent these types of failures Figure 4 BOV trend of a sensor exposed to long term excessive temperatures Erratic bias and time waveform The b...

Page 5: ...ply voltage and ensure that the bias voltage is centered between supply voltage and ground voltage However the bias voltage and power supply are rarely adjustable For ex ample if you are using an 18 v...

Page 6: ...cally coupled signals Using shielded twisted pair cable will also help minimize any magnetically coupled noise into the signal cable For example if a power cable is 440 volts and the vibration signals...

Page 7: ...am release air leak cavitation etc Repair steam leak dump Use less sensitive sensor Place rubber pad under sensor 10 0 14 0 V Stable Very high low frequency ski slope No high frequency signal Choppy D...

Reviews: