Select
‘
Yes
’
to change the default gauge labels.
Each gauge can have a user-defined, three-character gauge
label. Select each gauge in turn and change each character
with
△
or
▽
and confirm with ENT.
Relay assignment
You can assign any
relay to any gauge or to ‘None’. Relays
assigned to ‘None’ are permanently de
-energised.
Trip levels are set to maximum on assignment.
Trip levels for relays are set using the TRIPS push-button and
can be modified when the Function Switch is at position IG1,
OFF or IG2. Refer to section 7.2 for instructions on trip settings.
Relay D can also be assigned to Bake.
Bake
Set the bake temperature setpoint (50 to 250°C), total bake
time (1 Hr to 90 Hrs) and overpressure setpoint as required.
During Bake, relay D is energised when both the temperature
and pressure are below their setpoints. The hysteresis on
temperature is 5°C and on pressure is 50%. Refer to section 10
for more information.