www.amico.com 4
WARNING:
Unplug power supply from wall socket prior to cleaning the diagnostic station.
WARNING:
DO NOT autoclave the diagnostic station.
WARNING:
Liquids can damage electronics inside the diagnostic station. Prevent liquids from
spilling on or dripping into the diagnostic station.
Clean the diagnostic station on a routine basis according to your facility’s protocols and standards or
local regulations.
To clean the diagnostic station (excluding the instrument heads), use any of the following:
• A clean cloth slightly dampened with 70 percent isopropyl alcohol or 10 percent chlorine
bleach solution. Wipe dry with a clean cloth.
• Sani-Cloth Plus.
• CaviWipes.
Cleaning
Power supply:
Model #: RR9LA4000CCPYIM
Input: 100-240V, 50-60Hz, 1.5A
Output: 5V, 4A
Diagnostic Station:
Input: 5V, 4A
Electrical Specifications
Environmental Specifications
Operating:
Temperature limitation Humidity limitation Atmospheric pressure limitation
Transport/Storage:
Temperature limitation Humidity limitation Atmospheric pressure limitation
1060 hPa
1060 hPa
700 hPa
700 hPa
95%
95%
15%
15%
59 °F
(15 °C)
104 °F
(40 °C)
-4 °F
(-20 °C)
120 °F
(50 °C)
Approvals
UL 60601, CSA C22.2 No. 601.1