h. Ramp to 100.0 Amps via the MAXUS.
i. Adjust GAIN potentiometer P1 until the millivoltmeter reads 100.0 mV.
j. Adjust ANALOG OUT potentiometer P4 until the computer via MAXUS reads 100.0
Amps.
k. Adjust DVM OUT potentiometer P5 until the programmer front panel current meter
reads 100.0 Amps.
l. Repeat steps a. through l. until no further adjustments are necessary.
m. Place the front panel COMPUTER/MANUAL switch in the MANUAL position.
43. Place the front panel CURRENT CONTROL switch in the UP position.
44. At maximum current, adjust CURRENT LIMIT potentiometer P12 until the
millivoltmeter reads 100.0 mV.
45. Deenergize the system and reinstall the cover.
The calibration procedure is complete.