4-22 | IPS-4 Dual Bench (UV / IR) Analyzer
Inject Zero fluid
This function injects Zero fluid into the sample system to flush the
Measuring Cell of light-absorbing compounds. During this operation,
the zero solenoid is open and “Zero Hold” is displayed in the
Status
area of the
HOME
screen. Note that this mode does not actually Zero
the analyzer.
Inject Span fluid
This function injects Span fluid into the sample system, which can
be used to observe the analyzer’s response to species of known
concentration(s). During this operation, the span solenoid is open and
“Span Hold” is displayed in the
Status
area of the
HOME
screen. Note
that this mode does not actually Span the analyzer.
Start measure
This function returns the analyzer to normal
Measure
mode from any
of the
Inject Zero fluid
,
Inject Span fluid
,
Zero Gas Flush
, or
Span
Gas Flush
functions. To return the analyzer to
Measure
mode, select
this function and press
ENTER
. The message
“Please wait .
..” appears
while the analyzer performs a Sample Flush before switching back to
Measure
mode.
Manually Starting a Zero Calibration
Refer to “Manually Zeroing the Analyzer Sample System” in Chapter 3 for
information about how to perform a manual zero on the analyzer.
Summary of Contents for Western Research IPS-4
Page 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Page 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Page 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Page 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Page 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Page 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Page 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Page 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Page 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Page 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Page 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Page 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Page 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Page 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Page 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Page 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...