2-4 | Thermox WDG-V / VC / VCM with Blow Back
For measuring oxygen in non-combustibles gases, the calibration of an
analyzer is obtained from the formula:
AT = 48.0 at 680°C
E = A*T*Log
20.9%
O
2
Unk%
Where A is a constant, T is the cell temperature on an absolute scale (°C +
273) and O
2
Unk% is the unknown oxygen concentration of the gas to be
analyzed, and which is calculated by the analyzer.
The cell produces zero voltage when the same amount of oxygen is on
both sides, and the voltage increases as the oxygen concentration of the
sample decreases. The voltage created by the difference in the sample gas
and the reference air is carried by a cable to the microprocessor control
unit, where it is linearized to an output signal.
Because of the high operating temperature of the cell, combustible
gases that are present may burn. When this occurs, the cell will gener-
ate high millivolts and cause the display to indicate less oxygen than
is actually in the gas (net oxygen content).
Hydrocarbons
When hydrocarbons are present in the gas sample, a combustion process
occurs when this gas sample is exposed to the high temperature of the
zirconia cell. An indication that hydrocarbons may be present in the gas
sample is that the oxygen reading will be lower than expected. For exam-
ple, if a calibration gas cylinder has an oxygen value of 20 PPM and 5 PPM
of hydrogen (balance nitrogen), the oxygen analyzer will read 17.5 PPM
oxygen. The reduction of oxygen is due to the combustion process where
5 PPM of hydrogen will combine with 2.5 PPM of oxygen to form water.
Thus, the oxygen analyzer will read 17.5 PPM rather than the actual 20
PPM oxygen.
The amount of oxygen reduction is dependent on the type of hydro-
carbons present in the sample gas.
NOTE
NOTE
Summary of Contents for Thermox WDG-V
Page 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Page 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Page 73: ...AMEVision Display User Interface or Communications 4 5...
Page 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Page 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Page 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Page 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Page 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Page 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Page 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Page 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...