Installation Guide | 3-25
Standard Connection using PC Configurator Software
If you are connecting the analyzer using the AMEVision Display
User Interface, skip this section and continue with the AMEVision
Display Installation Guide following this section.
Overview
The WDG-V UOP PC Configurator Software comes standard with the
analyzer and is used to configure, calibrate and monitor a single analyzer.
It is connected to the PC using a two-wire MODBUS RTU interface.
Connecting to the WDG-V UOP
Setting the Address using the Dip Switch
The address on the WDG-V UOP analyzer is set using the dip switch on
the front of the electronics PCB. When connecting a single analyzer, the
address is “1”.
NOTE:
ORIENTATION OF SWITCH MOUNTED IN ANALYZER IS UPSIDE DOWN.
Figure 3-11. Close up of dip switch.
321 Switch
Address
001
1
1 = ON (UP)
0 = OFF (DOWN)
The AMEVision Display User Interface and PC software cannot be
connected to the analyzer at the same time.