Installation
2-10
TM-F1OP-C1XN-01
Functional Tests
These functional test procedures include power-on and front panel
function checks as well as voltage and current mode operation checks.
Equipment Required
•
Digital voltmeter (DVM) rated better than 0.5% accuracy.
•
DC shunt 1mV/A (±0.25%) with connecting wire. The recommended
current ratings for the DC shunt and the wire must be at least 10%
more than the output current of the power supply.
Power-on Check
To perform the power-on check:
1. Ensure that the AC power switch is in the OFF position.
2. Ensure that the output sense lines are connected in the default
configuration. (The local sense lines are connected between
terminals 1 and 2 and between terminals 4 and 5 on the J10 sense
connector as shown on the rear panel diagram in
3. Turn the voltage and current controls fully counter-clockwise.
4. Connect the unit to an AC outlet.
5. Turn the front panel AC power switch to ON.
After a short power-on delay, the front panel digital meters light up and
the green voltage mode LED turns on. Both voltmeter and ammeter
displays read zero.
TM-F1OP-C1XN-01.book Page 10 Thursday, October 25, 2007 9:04 PM
Summary of Contents for Sorensen XFR 1200 Watt Series
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...This page intentionally left blank ii...
Page 8: ...This page intentionally left blank iv...
Page 10: ...This page intentionally left blank vi...
Page 16: ...xii TM F1OP C1XN 01 book Page xii Thursday October 25 2007 9 04 PM...
Page 18: ...xiv TM F1OP C1XN 01 book Page xiv Thursday October 25 2007 9 04 PM...
Page 30: ...1 12 TM F1OP C1XN 01 book Page 12 Thursday October 25 2007 9 04 PM...
Page 58: ...2 28 TM F1OP C1XN 01 book Page 28 Thursday October 25 2007 9 04 PM...
Page 82: ...4 8 TM F1OP C1XN 01 book Page 8 Thursday October 25 2007 9 04 PM...
Page 94: ...5 12 TM F1OP C1XN 01 book Page 12 Thursday October 25 2007 9 04 PM...