background image

XCEL

 255 

Slit Lamp

 

User’s Guide

Summary of Contents for Reichert XCEL 255

Page 1: ...XCEL 255 Slit Lamp User s Guide...

Page 2: ...s the right to make changes in the product described in this manual without notice and without incorporating those changes in any products already sold ISO 9001 13485 Certified Reichert products are d...

Page 3: ...Input Power 11 Disconnection of Input Power 11 Instructions for Use 12 Operation 12 Slit Length 13 Filter Dial 13 Slit Rotation 13 Cleaning Maintenance 14 External Cleaning 14 Forehead Chinrest Prepa...

Page 4: ...st only be connected to a supply mains with protective earth OR DAMAGE TO THis instrument AND OR INJURY TO THE OPERATOR OR PATIENT MAY OCCUR WARNING ensure that the voltage applied to the unit is the...

Page 5: ...AY OCCUR SEE MAINTENANCE SECTION FOR DETAILED CLEANING INSTRUCTION CAUTION MEDICAL ELECTRONIC EQUIPMENT NEEDS SPECIAL PRECAUTIONS REGARDING EMC AND NEEDS TO BE INSTALLED AND PUT INTO SERVICE ACCORDING...

Page 6: ...rrent Power Protective Earth Connection ON OFF 2013 Date of Manufacture REF Catalog Number S N Serial Number Waste of Electrical and Electronic Equipment Compliance to Medical Device Directive 93 42 E...

Page 7: ...should operate this instrument Please retain this manual for future reference and to share with other users Additional copies can be obtained from your authorized Reichert dealer or from the Reichert...

Page 8: ...Rotation Scale 14 Slit Width Rotation Knob 15 Illumination Arm Lock Knob 16 Microscope Arm Lock Knob 17 Instrument Base Lock Knob 18 Guide Rail Covers 19 Geared Rollers 20 Guide Rails 21 Chin Rest Ass...

Page 9: ...Screw provided torque to 7 8 N m Refer to Figure SU 3 Using the 5mm Hex Wrench remove the two 7 Screws from the bottom of the Table Top and attach the Chin Rest Assembly to the Table Top using these S...

Page 10: ...8 Note There is a Notch in the Base Assembly and a Slot in the Illumination Assembly Align the Illumination Assembly and Arm so the Notch goes into the Slot If the Notch is not aligned properly the Sl...

Page 11: ...TE THE DISCONNECTION DEVICE PLUG After the unit is in its secure location apply the 1 correct input voltage to the instrument using the Power Cord from the Accessory Tray Note The power inlet is locat...

Page 12: ...post Refer to Figure IN 2 Microscope elevation is adjusted by rotating the 7 Joystick and observing the slit image through the Microscope Assembly until the slit is centered on the patient s cornea Re...

Page 13: ...y index into place Refer to Figure IN 7 Filter Dial The Filter Dial has four positions that index into place and are color coded to indicate the active filter Refer to Figure IN 7 The color coded inde...

Page 14: ...e a bulb that has recently been in use as it will be very hot Wait until it has cooled and use gloves or a thick cloth when handling any halogen bulb WARNING Never touch a Halogen Bulb with bare hands...

Page 15: ...tly pull out the Fuse Holder by gripping the two small tabs Refer to Figures MM 5 and MM 6 Open the Door to the Fuse Holder by pulling it 3 down Refer to Figure MM 6 Note The Fuses will pop up when th...

Page 16: ...ion Light Harness not plugged into the Power Supply Assembly Ensure the Fixation Light Harness is properly seated in the Power Supply Assembly Defective Power Supply Replace the Power Supply Light too...

Page 17: ...lative Humidity 30 to 75 Atmospheric Pressure 80 kPa 23 6 in Hg to 106 kPa 31 3 in Hg Transportation Storage Temperature 20 C 4 F to 70 C 158 F Relative Humidity 10 to 80 non condensing Atmospheric Pr...

Page 18: ...0 16 465 mm x 316 mm Disposal This product does not generate any environmentally hazardous residues At the end of its product life follow your local laws and ordinances regarding the proper disposal o...

Page 19: ...nment Emissions Test Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 Class B The Xcel 255 uses RF energy only for its internal func tion Therefore its RF emissions are ve...

Page 20: ...hospital environment Surge EN IEC 61000 4 5 1kV Differential 2kV Common 1kV Differential 2kV Common Mains power quality should be that of a typical commercial or hospital environment Voltage Dips Drop...

Page 21: ...trengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance levels in each frequence range V1 and E1 Interference may occur in the vicinity o...

Page 22: ...imum output power of the communications equipment Max Output Power of Transmitter W Separation m 150kHz to 80 MHz d 3 5 V1 Sqrt P Separation m 80 to 800 MHz d 3 5 E1 Sqrt P Separation m 800MHz to 2 5G...

Page 23: ...RRANTY If notified promptly in writing of any action brought against the purchaser based on a claim that the instrument infringes a U S Patent Reichert will defend such action at its expense and will...

Page 24: ...15120 101 Rev D March 12 2013...