3. MEASUREMENT OF SMALL CURRENT SIGNALS 

This chapter describes the issues surrounding the measurement of small current signals to include proper 
shielding and cell design recommendations. 
 

3.1. Faraday Shield 

The LCI and corresponding potentiostat system operate near the limits of current measurement. The LCI 
system can resolve current in the attoamp (10-18) range. For perspective, 100aA represents the flow of 
around 600 electrons per second! 
 
A Faraday shield for low current measurements is mandatory. A Faraday shield is a conductive enclosure 
that  surrounds  the  cell.  The  materials  can  vary  from  a  metal  box  or  fine  wire  mesh,  to  a  plastic  case 
covered  in  conductive  paint.  The  conductive  coating  should  be  continuous  on  all  sides  of  the  cell 
(including the bottom), and all parts operate as a single, conductive unit. 
 
For the LCI, the best shielding will occur if the LCI main body and the cell leads are all placed inside the 
Faraday  shield,  as  the  shield  must  be  connected  to  the  LCI  ground.  Note  the  photos  below  of  the 
Princeton Applied Research K0269B Faraday box. The box is made of steel with a few small openings for 
cell  cables  or  other  leads.  The  box  can  be  used  in  a  vertical  or  horizontal  position.  In  the  photo  to  the 
right, the LCI Main body has been attached to the wall of the K0269B using  Velcro tape (not supplied). 
Also note in the upper left corner above the cell is a grounding point to connect the ground lead from the 
LCI to the Faraday shield (not connected in photo).  
 

 

 

 

Figure 15. K0269B Faraday Box 

 

 

NOTE 

The ring-stand and cell are not included with the K0269B Faraday box. 

 

Summary of Contents for PARSTAT 4000

Page 1: ...VersaSTAT LC Low Current Interface LCI Hardware Manual P N 224186 Rev B...

Page 2: ...LCI Main Body 5 2 1 2 Connect Interface Cable to Potentiostat Cell Interface 5 2 1 3 Connect Cell Cables to LCI Front Panel 5 2 2 LCI Calibration 5 2 2 1 Calibration 5 2 3 LCI Checkout Procedure 9 3 M...

Page 3: ...4 LCI with Cell Cables Connected Across Internal Feedback Dummy Cell Resistor 3 Figure 5 Cell Cables 4 Figure 6 Calibration Required Message 6 Figure 7 Calibration Procedure 6 Figure 8 Calibrate Mess...

Page 4: ...Indicates Warning or Caution Indicates Danger due to Electric Shock Hazard Please read all safety instructions carefully and make sure you understand them fully before attempting to use this product C...

Page 5: ...e refer to the VersaStudio operation manual Chapter 2 describes how to properly install calibrate and test the LCI option Chapter 3 discusses the measurement of small current signals with recommendati...

Page 6: ...l cell leads that connect from the LCI main body to the electrodes and or cell under test as well as a ground lead 1 2 1 LCI Interface Cable The LCI potentiostat interface cable part number 223763 sho...

Page 7: ...3 LCI Cell Cable Panel Left and Potentiostat Interface Panel Right Figure 4 LCI with Cell Cables Connected Across Internal Feedback Dummy Cell Resistor The WE cable connects to the W pin jack the SE c...

Page 8: ...enance There are no user serviceable parts in this instrument Breaking the seal by opening the cover will void your warranty Contact the factory service department or the affiliate in your area if you...

Page 9: ...WE green SE gray RE white and CE red 2 2 LCI Calibration The VersaSTAT LC Low Current Interface is designed to provide ultra low current measurements beyond the capabilities and specifications of the...

Page 10: ...brate Now button Figure 7 Calibration Procedure NOTE It is very important that the calibration be performed in the exact same manner as the experimental settings The lowest current ranges and current...

Page 11: ...Figure 8 Calibrate Message The configuration in Figure 9 shows where the WE SE and RE are connected but the CE is not Once connected select OK to continue Figure 9 CE Not Connected After a few more m...

Page 12: ...also connected between the LCI and K0269B to properly ground the Faraday box to the same chassis ground as the LCI and VersaSTAT Figure 12 LCI Inside Faraday Box After completing the calibration the...

Page 13: ...ting the Line Fit button at the top of the graph 4 The vertex currents should be 10nA and the slope should be 1 0E 7 the inverse of which is 10Mohm The PARSTAT4000 VersaSTAT LC combination accesses ad...

Page 14: ...tion has finished select all of the data points Data Select All and perform a Line Fit by selecting the Line Fit button at the top of the graph 4 The vertex currents should be 10nA and the slope shoul...

Page 15: ...ating should be continuous on all sides of the cell including the bottom and all parts operate as a single conductive unit For the LCI the best shielding will occur if the LCI main body and the cell l...

Page 16: ...rements between electrodes at impedances of 1012 Ohms 3 3 3 Reference Electrode One of the most common problems with cell noise in electrochemical experiments in general is high impedance reference el...

Page 17: ...n still occur in some experimental conditions and are most often manifested as excessive noise or jumps in the data and graphical output If an oscillation is suspected one can use the LCI Bandwidth Li...

Page 18: ...VersaSTAT LC Low Current Interface LCI Hardware Manual P N 224186_0B 14 THIS PAGE INTENTIONALLY LEFT BLANK...

Page 19: ...um Input Voltage Differential 10 V Bandwidth 700kHz 3 dB Common Mode Rejection 60dB 100Hz 50dB 100kHz Input Impedance 1014 in parallel with 200 fF typical 4 1 4 Current Measurement Ranges 12 decades 2...

Page 20: ...ts of the following standards LVD EN61010 1 1993 Amendment 2 EMC EN61326 1998 Emissions EN55011 1991 Group 1 Class A Immunity IEC 61000 4 2 1995 ESD IEC 61000 4 3 1995 EM field IEC 61000 4 4 1995 Burs...

Page 21: ...ABILITY AND FITNESS AS WELL AS ANY AND ALL OTHER OBLIGATIONS OR LIABILITIES OF PRINCETON APPLIED RESEARCH INCLUDING BUT NOT LIMITED TO SPECIAL OR CONSEQUENTIAL DAMAGES NO PERSON FIRM OR CORPORATION IS...

Reviews: