
User Manual
–
Rev AV
AMETEK Programmable Power
MX Series
120
CLOCK/LOCK
Indicates the presence of the -LKS clock and lock auxiliary
option. For units without -LKS, this field is set to N/A. For units
with the -LKS option installed, this field is set to ON. Note that
master unit (with -LKM) also has this field set to N/A.
WH METER
Indicates the presence of the Watt Hour Meter option.
SYSTEM
Shown here on Series I models only. See CONFIGURATION
SETUP 3 screen below.
CONFIGURATION SETUP 3
(Series II models only)
MS704
Indicates the presence of the MIL/STD-704 Revision A through
F test option. If this option is installed, this field will show ON. If
this option is not installed, this field will show N/A (not available).
This field is available on Series II MX systems only.
ABD
Indicates the presence of the ABD0100.1.8 test option. If this
option is installed, this field will show ON. If this option is not
installed, this field will show N/A (not available). This field is
available on Series II MX systems only.
LF
Indicates the presence of the Low Frequency limit option. If this
option is set, the maximum frequency that can be programmed
is 500 Hz. This field is available on Series II MX systems only.
SYSTEM
This field sets the controller for the correct MX system
configuration. Available settings are:
MX30
MX30 Single Unit
MX45
MX45 Single Unit
MX60
MX60 system Master
MX90
MX90 system Master
MX135
MX135 System Master
MX180
MX180 system Master
MX225
MX225 system Master
MX270
MX270 System Master
AUX Auxiliary
System.
If this field is set incorrectly, the current limit scaling and current
measurement will be off by a factor of three. Changes made to
the SYSTEM field will not take effect until
AFTER
power on the
MX system is cycled. (Off, the back on).
Note: MX systems with Firmware rev. less than 4.0 do not
support MX60 configuration and uses MX45 setting for MX30.
MB
Indicates the presence of the Multi-box option. If this option is
set, the SYSTEM field (above) can be changed to
accommodate reconfiguration of the MX system. If this option is
not set, the SYSTEM configuration is fixed and cannot be
changed by the user. This field is available on Series II MX
systems only.
NOTE: If the MB option is enabled, great care MUST be taken to set the correct SYSTEM
setting for the configured hardware. Failing do so will result in incorrect operation
of the MX system and could even result in damage of MX or the equipment under
test.