Page 2: ...This blank page has been inserted to enable double sided printing of the document...

Page 3: ...Dansensor MAP Mix 9001 ME EN User Manual P N 250151 P 10 2021 UserManual Published by MOCON Europe A S R nnedevej 18 4100 Ringsted Denmark Tel 45 57 66 00 88 info dk mocon ametek com www dansensor com...

Page 4: ...2 User Manual EN Dansensor MAP Mix 9001 ME P N 250151 P 10 2021...

Page 5: ...on 5 2 Safety instructions 7 3 Equipment set up 9 Gas connection 9 Mounting of filters and gas connection fittings 11 Electrical connection 13 4 Control panel 15 Flow control 15 5 Spare parts and opti...

Page 6: ...4 User Manual EN Dansensor MAP Mix 9001 ME P N 250151 P 10 2021...

Page 7: ...industrial applications where the gases N2 CO2 and O2 are used in either 2 gas or 3 gas combinations The mixing principle of Dansensor MAP Mix 9001 ME is very stable and can be used for all types of...

Page 8: ...6 User Manual EN Dansensor MAP Mix 9001 ME P N 250151 P 10 2021...

Page 9: ...O2 oxygen Do NOT use this equipment for air supply for human beings When using the mains supply Dansensor MAP Mix 9001 ME must be grounded WARNING Always use fittings and hoses which apply to the gass...

Page 10: ...8 User Manual EN Dansensor MAP Mix 9001 ME P N 250151 P 10 2021...

Page 11: ...the existing assemblies see Fig 2 Fig 2 Correct mounting dismounting of fittings DANGER At the back of the mixer there are labels showing which gas to connect to the individual inlets see Fig 3 NEVER...

Page 12: ...nsensor MAP Mix 9001 ME P N 250151 P 10 2021 WARNING On buffer tank versions make sure that the gas outlet is NEVER blocked or restricted in front of the buffer tank see Fig 4 Fig 4 Correct placement...

Page 13: ...d and supplied last Models 400 liter min Mounted from the factory A Gas inlet Nylon gasket 1 4 x 1 5 mm Filter sinterbronze 60 O ring 13 94 x 2 62 mm Loctite 542 hydraulic Reduction nipple 1 2 3 8 Sup...

Page 14: ...oarse O ring 13 94 x 2 62 mm Loctite 542 hydraulic Reduction nipple 1 2 3 8 Supplied parts B Nylon gasket 3 8 x 1 5 mm Hose nipple 3 8 10 11mm Hose clamp 13 20 mm ATTENTION ALWAYS use Loctite 542 when...

Page 15: ...d in two different ways Between Green N O and Yellow COM relay is normally open and closes in case of an alarm Between Grey N C and Yellow COM relay is normally closed and opens in case of an alarm FL...

Page 16: ...14 User Manual EN Dansensor MAP Mix 9001 ME P N 250151 P 10 2021...

Page 17: ...3 lights up at too low pressure on gas GAS IN 3 gas mixer only The lamp F lights up when the solenoid valve for gas flow is open optional Flow control Turn the FLOW dial for setting of the wanted gas...

Page 18: ...7 5 5 The O2 gas constitutes the rest 10 Example 2 30 CO2 20 N2 50 O2 1 Set the CO2 dial at 30 2 Read the value of X 7 3 Calculate MIX N2 X 20 7 2 86 4 Set the MIX dial to 2 9 5 The O2 gas constitutes...

Page 19: ...on off button B 3 gas mixer used as 2 gas mixer How to use the mixer for only N2 and CO2 1 Adjust the CO2 on the CO2 dial 2 Never mind the MIX dial setting How to use the mixer for only N2 and O2 1 S...

Page 20: ...18 User Manual EN Dansensor MAP Mix 9001 ME P N 250151 P 10 2021...

Page 21: ...valve 24VDC 1 2 P N 250154 Kit pressure section all MAP Mix and TGC ser cpl P N 251382 Pressure regulator 0 7 bar O2 cleaned P N 270369 Pressure switch 1 10 bar n closed O2 cleaned P N 270383 Pressure...

Page 22: ...20 User Manual EN Dansensor MAP Mix 9001 ME P N 250151 P 10 2021...

Page 23: ...6 400 2 100 400 5 5 10 3 8 6 200 3 20 200 5 5 10 2 8 5 400 3 100 400 5 5 10 2 8 5 Type Gasses Capacity l min Inlet pressure bar 250 2 25 250 5 5 10 400 2 100 400 5 5 10 200 3 20 200 5 5 10 400 3 100 4...

Page 24: ...bar Max 10 bar The specific pressure interval is indicated on the mixer Outlet pressure 2 gas 3 8 6 bar 55 87 psi 3 gas 2 8 5 bar 41 73 psi Gas connections Models 400 liter min 3 8 RG female Fittings...

Page 25: ...active Signal cable Alarm output N O N C COM connection Max 48V 1A Flow control valve Option Connection of 24 VDC Material Stainless steel IP 54 Ambient temperature Operational 0 to 45 C 95 RH non con...

Page 26: ...24 User Manual EN Dansensor MAP Mix 9001 ME P N 250151 P 10 2021...

Page 27: ...This blank page has been inserted to enable double sided printing of the document...

Page 28: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...

Reviews: