![Ametek Mocon Dansensor LeakPointer 3 User Manual Download Page 85](http://html1.mh-extra.com/html/ametek/mocon-dansensor-leakpointer-3/mocon-dansensor-leakpointer-3_user-manual_2931374085.webp)
Dansensor
®
LeakPointer 3/3
+
EN
User Guide
83
Network Logging
The
Network Logging
screen holds the parameters for the network logging function.
If enabled, the function sends a data string through a LAN connection for each measurement.
NOTE! Always consult your network administrator before making network
settings, as incorrect settings can result in reduced or no network activity.
A TCP/IP LAN server has to be implemented to receive data. The server will listen to the port as
specified above.
The protocol is an array of semicolon separated values:
"Device Serial No; Date; Time; User Name; Product Name; Product ID; Mode; Standard Test;
State; Vacuum (delta); Test time; Leak limit; Hole limit; Leak gain factor; Hole size (micron); CO2
increase /ppm/s); Leak; Marked state; Error nr; Days till calibration; Invalid measurement
comment; Custom field 1; Custom field 1 reply; Custom field 2; Custom field 2 reply; Custom
field 3; Custom field 3 reply; Custom field 4; Custom field 4 reply;"
Network Logging
Select whether or not network logging is required.
Disabled = No network logging
Enabling
Network Logging
also enables the following
parameters:
Server IP
Set up an IP-address to be used for
collecting log data for each measurement
via LAN.
This requires for setting up of a
Server Port
number as well.
Server Port
See above.
Acknowledge
Byte sent by the server to the device after
each received log entry.
Setting value to 0 will disable the function.
Summary of Contents for Mocon Dansensor LeakPointer 3
Page 1: ...P N 380064 F 06 2021 UserGuide...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 8: ...6 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 20: ...18 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 94: ...92 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 98: ...96 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 99: ...This blank page has been inserted to enable double sided printing of the document...
Page 100: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...