Dansensor
®
LeakPointer 3/3
+
EN
User Guide
53
8. When the measurement is complete, the result is displayed on the screen.
Results for
Single Package
mode measurements are read out as the calculated hole size
in
μm
, while
Multi Package
mode results are read out as the measured increase in the
CO
2
level in the chamber in
ppm/s
.
The
symbol means that the measurement is within the allowed range, while the
symbol means that the measurement is higher than the allowed range.
The small statistics counter in the lower left corner of the display will be updated
accordingly.
9. If
Acknowledge Measurement
is enabled in
Main Menu -> Workflow
you must
press the
,
, or
button now to validate and end the measurement/workflow - see
for details.
10. The device is now ready for a new measurement/workflow.
Summary of Contents for Mocon Dansensor LeakPointer 3
Page 1: ...P N 380064 F 06 2021 UserGuide...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 8: ...6 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 20: ...18 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 94: ...92 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 98: ...96 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 99: ...This blank page has been inserted to enable double sided printing of the document...
Page 100: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...