If the ISM-3 is equipped with an internal pump, the gas is sucked from measure point at
atmospheric pressure. When calibrating, the gas is sucked from a gas overflow, to which
approx. 0.5 l/min must be supplied.
Fig. 6. Flow diagram of ISM-3 with pump. Gas overflow for calibration use is shown
Operation and connection
Operation of ISM-3:
Display
4 digit red LED display
Control
4 front keys. Key function is activated when key is released.
Measuring accuracy
Better than +/- 1% of the measured value
+/- 1 digit in calibrated measuring range
Measuring range
20.9% - 1ppm (standard) *
* For measuring in other ranges device must be recalibrated - see
for details
Possible connection to the ISM-3:
Gas supply
Gas inlet on the back
Relay
Max. 48V, 1A. (COMMON, N.O. and N.C.)
Current output
Programmable 0-20 or 4-20 mA, with user-defined scale
(e.g. 0-1%, 10-20.9% or 0-100ppm O
2
).
Voltage output (Option)
Programmable 0-10 volt or 2-10 volt with user defined scale
Measure input
10-32VDC external pump control. Consumption 10mA.
RS232C (Option)
For serial communication with a PC (for internal use only)
All electrical inputs and outputs, except for the RS232C port, is galvanically separated from
the internal electronics by means of optocouplers and/or a relay.
(for flow indication)
Restriction hose
Gas out
125 ml/min
Pump
Gas in
0,5 l/min
Hoses:
ID = min. 3 mm
L = min. 0.5 m
O
2
sensor
Pressure
sensor
Summary of Contents for mocon Dansensor ISM-3
Page 1: ...P N 200114 I 09 2021 UserManual...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 6: ...4 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 10: ...8 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 12: ...10 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 20: ...18 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 28: ...26 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 32: ...30 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 33: ...This blank page has been inserted to enable double sided printing of the document...
Page 34: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...