AMETEK CTS
H-Field Test
Operating Manual
V 1.20
28 / 37
2.5.1.
Quick Start
Easy and very fast operation of all standard functions of the equipment. The latest simulator settings are stored
automatically and will be recalled when Quick Start is next selected.
Quick Start Menu
Quick Start testing for 100 A/m
2.5.2.
Standard test Routine
The user can select preprogrammed standard test routines.
Change to
- Quick Start
- User Test
Other:
- Add to STEPS
- Configuration
- Save as User
- Load Defaults
Change to
Standard
library
Within this test routine all standard parameters can be changed online during testing. For change tip to the black
marked parameters
The user can easy select the following parameters
T
(test duration):
1 s to 999 s (standard says
60 s
or faster depend on the protection devices power rate)
Level
(voltage):
Level 3 (100 A/m)
Level 4 (300 A/m)
Level 5 (1000 A/m)
Coupling
(Cpl):
HV - COM
(to antenna)