H-Field Test
Operating Manual
V 1.20
20 / 37
2.2.
IEC 61000-4-8 50/60 Hz magnetic field with NSG 3040A/3060A
The Power magnetic menu offers different test routines for magnetic field testing.
Click into Power Magnetic Field
menu loads the
last used menu
of the following routines below.
Quick Start
Easy and fast online-operation with the phenomenon magnetic field. In this menu the user can operate a
magnetic field test manually with online change of most parameters during a test.
Standard
The operator can select between various preprogrammed test routines as required in different standards
The standard library is filtered according the EUT setting (AC, DC etc.)
User test routines
The user defined library where all created magnetic field tests are stored.
2.2.1.
Quick Start
Easy and very fast operation of all standard functions of the equipment. The latest simulator settings are stored
automatically and will be recalled when Quick Start is selected.