103887 01
9
Press
to save the present selection.
3.6.3 Selection of STEP SIZE
After SLOPE TIME has been selected, STEP SIZE (step size in % of range) can be
selected for the transmitter types LIN and FLOW. If MODE 4-20mA VALVE is selected, a
fixed STEP SIZE is applied. The text
is shown in the display for approx. 1 second.
Press
or
for setting of STEP SIZE. A choice can be
made between the step sizes 5%, 10%, 20%, 25%, 50% and 100%.
Press
to save the present selection.
3.6.4 Selection of STEP TIME
After STEP SIZE has been selected, STEP TIME (the time between changes in the
AUTO-STEP function) can be changed. Either
s
or
s
is shown in the display.
Press
or
for setting of STEP TIME.
Press
to save the present selection.
3.6.5 Selection or de-selection of automatic OFF function
After STEP TIME has been selected, it is possible to select or de-select the AUTOMATIC
OFF function (mAcal automatically switched off approx. 30 minutes after a key has last
been activated).
The text
is shown in the display for approx. 1 second. Alongside, the text
or
appears.
Press
and AUTOMATIC OFF is selected.
appears in the
display.
Press
and AUTOMATIC OFF is deselected.
appears in the
display.
Press
to save the present selection.
SETUP is exited and the mAcal is now ready for normal operation with new parameters.