5  Commissioning

5.1  Inspection on receipt

Physically examine all items for damage that may have occurred during transit.  

Check the contents against the packing note.
If any item has been damaged in transit, this must be reported to the carrier 

and to the supplier immediately.  Do not return damaged instruments to the 

sender as the carrier will not then consider a claim.  Save the packing with the 

damaged article for inspection by the carrier.

5.2  Connections to the electrical supply

The a.c. supply cable has three cores.  Connect the brown lead to live, blue to 

neutral and green/yellow to earth.  When connected to the electrical supply and 

switched on the instruments will light.

Warning

At no time must any panels be removed when connected to the supply.

5.3  Heating up of the source

         Caution 

Ensure that there is free space around the source.  Do not obstruct any 

of the vents in the control section, they are needed to keep the controls 

cool.  When the source is operating at any temperatures above ambient, 

the front plate and case will become hot.

The source can be operated at any temperature in the range ambient to 1200°C 

(2200°F), however the recommended temperature range is 350°C to 1150°C 

(650°F to 2100°F).  Set the controller to the required value as follows:-
1)  When first switched on, the controller goes through a self check procedure.  

When this check is completed two displays will be visible.
Upper display   

Actual element temperature

Lower display   

Set point temperature

2)  To modify the set point press, the ‘Up’ or ‘Down’ button until the required 

value is obtained.
The source will be ready for use at 1100°C/2000°F approximately 20 

minutes after switch on.
The output light ‘OP’ indicates the state of the controllers output signal 

to the solid state relay which in turn switches the power to the heating 

element.

Note

All other controller parameters are factory set and locked.  It should not 

be necessary to adjust them.

Summary of Contents for Landcal R1200P

Page 1: ...le 01246 410585 Email land infrared ametek co uk Internet www landinst com Land Instruments International 2008 2012 AMETEK Land Inc 150 Freeport Road Pittsburgh PA 15238 U S A Telephone 412 826 4444 F...

Page 2: ...at safety precautions are taken whenever the materials are handled Exposure to fibrous dust may cause respiratory disease When handling the material always use an approved respiratory protection equip...

Page 3: ...n It is possible that this equipment is to be installed on or near to machinery or equipment operating at high temperatures and high pressures Suitable protective clothing along with face and eye prot...

Page 4: ...s etc Design and Manufacturing Standards The Quality Management System of Land Instruments International is approved to BS EN ISO 9001 for the design manufacture and on site servicing of combustion en...

Page 5: ...Source Type R1200P 1 2 Description 3 3 Specification 4 4 Electrical Supply details 6 5 Commissioning 7 6 Using the Furnace 9 7 Calibration of Radiation Thermometers 11 8 Maintenance 12 9 Eurotherm Te...

Page 6: ...Landcal Blackbody Source Type R1200P User Guide Page 6...

Page 7: ...the thermometer can be conducted by using the target indication From previous work the target indication temperature has been found to agree with the radiance temperature to within 10 C at short wave...

Page 8: ...some individuals be an irritant to the skin nose and throat Safety Note Refractory Fibrous Insulation This product contains alumino silicate wool products in its thermal insulation These materials ma...

Page 9: ...Fig 1 Landcal Blackbody Source Type R1200P CA970259 If the source is used as a transfer source i e calibration by comparison method adopted a set of optional standard radiation thermometers will be re...

Page 10: ...roximately 0 98 at short wavelengths Themocouple type Both the Control thermocouple and Indicator thermocouple are Type N NiSil NiCrNiSil mineral insulated 3mm OD with Nicrobell B sheath 350mm long pl...

Page 11: ...y R Red Element TC N L Indicator Controller Voltage change over switch shown in 220 240 position Panel Mounted Fuse 10A Super Quick Blow FF BU BU BU BU R R R GR Y Green Yellow R R PI G GR Y 4 6 2 3 5...

Page 12: ...nnected from the supply The isolator or the plug must be within easy reach of the operator The supply MUST incorporate an earth ground The wiring diagram for the R1200P is given in Fig 2 The power cab...

Page 13: ...the source Do not obstruct any of the vents in the control section they are needed to keep the controls cool When the source is operating at any temperatures above ambient the front plate and case wil...

Page 14: ...n or 25 pin to connect to the computer RS232 Cable Furnace to PC connections Furnace end of cable female Computer end of cable male 9 pin 9 pin 25 pin Rx 3 3 2 Tx Tx 2 2 3 Rx Com5 5 7 Com 7 8 4 5 Link...

Page 15: ...curve can be obtained with a Certificate of Calibration for the source A UKAS United Kingdom Accreditation Service National Measurement Accreditation Service Certificate of Calibration is available fr...

Page 16: ...e and carry the source in its accessory carrying case 3 After use it is recommended that the source be left to cool to an internal temperature of 200 C before moving 4 Never place the source into the...

Page 17: ...ing temperature of the thermometer to be tested The target temperature is that indicated by the measuring thermocouple displayed on the target indicator 7 2 Thermometer calibration When the source is...

Page 18: ...f the heating element and insulation 8 2 Fitting a replacement heating element insulation package The location of the terminal blocks on the insulation assembly is shown in Fig 3a The replacement asse...

Page 19: ...which terminal 7 Remove the insulation assembly from the base 4 set screws on the underside complete with earth connection cable To fit a new insulation assembly 1 Attach the insulation assembly onto...

Page 20: ...ff white paste to give good heat transfer to the aluminium sheet New relays are supplied with a heatsink pad All remnants of the old heatsink paste must be removed The thin white heatsink pad must be...

Page 21: ...trols when the set point is reached 8 5 Storage and carrying case A custom built aluminium storage and transportation case is available as an optional extra Use of this case is recommended when a The...

Page 22: ...res contain factory set parameters which should not be changed The Up and Down keys are used to alter the setpoint temperature in Level 1 operation and parameter values when in Level 2 operation To en...

Page 23: ...ramp rate allowing heating and cooling at the maximum rate OFF is the default setting 9 4 Altering the Power Limit when applicable 1 Press the Scroll key until OP Hi Output High is displayed 2 Press...

Page 24: ...Spare control or measuring thermocouple 135 172 11 Accessories The following calibration accessories are available for use with the Landcal type R1200P Description Land Part Number Storage carrying c...

Page 25: ...ters accessories and mountings and special instruments based on UNO Application dedicated processors based on LANDMARK Graphic ABTS S and ABTS U FTS VDT S and VDT U DTT FLT5 A This 36 month warranty i...

Page 26: ...onditions outside the Ambient Temperature Humidity and Vibration rating of the product environmental contamination solvent vapours deposition of airborne contamination cooling liquids of non neutral p...

Reviews: