8.0 Specifications
All specifications cover the temperature range from 0°C to +50°C, unless
otherwise noted.
Available Input Ranges
See page 13 for a table of available ranges in PSI plus equivalent ranges
and resolution for all engineering units
Accuracy
Positive Pressure: ±0.05% FS
Vacuum:
100, 300, 500 PSI indicator / 7, 21, 35 bar indicator: ±0.25% F.S.
Vacuum:
30 PSI / 2 bar indicator: ±0.1% F.S.
Over Pressure Protection:
Ranges from 15 PSI to 500 PSI/1 to 35 bar: 3X input pressure range
1,000, 3,000, and 5,000 PSI/70 to 350 bar ranges: 2X input pressure range
10,000 PSI/700 bar: 1.5X input pressure range
Overload Alarm (indicate OL on display): 1.2X input pressure range
Temperature Compensation
Full temperature compensated:
0 °C to +50 °C (32 °F to +122 °F) to rated accuracy
Note: For temperatures from –10 °C to 0 °C and 50 °C to 55 °C add .005%
F.S./°C
Standard Engineering Units
PSI, Bar, kg/cm2, inH
2
O (4 °C, 20 °C or 60 °F),
ftH
2
O (4 °C, 20 °C or 60 °F), cmH
2
O (4 °C and 20 °C), mH
2
O (4 °C and 20
°C), KPa, mBAR, inHg, mmHg, Torr
One custom unit (user programmable)
Media Compatibility
Liquids and gases compatible with 316 stainless steel
Environmental
Operating Temperature: -10 °C to +55 °C
Storage: -20 °C to +70 °C (-4 °F to +158 °F)