9.4 Calibration of analog out (DTI 1000 only)
Use following procedure for recalibration of the two analog outputs:
1.
Select the menu analog out.
2.
Enter the date format as YYMMDD.
3.
Connect the voltmeter to channel 1 as shown in the figure section 5.2.2. Read the
voltage and enter the measured value for channel 1.
4.
Connect the voltmeter to channel 2 as shown in the figure section 5.2.2. Read the
voltage and enter the measured value for channel 2.
Lowest possible voltages
Enter - Analog output channel 1 :
-0.6
Volt
Enter - Analog output channel 2 :
-0.6
Volt
< ESC > aborts
5.
Connect the voltmeter to channel 1 again, and enter the measured signal for
channel 1.
6.
Connect the voltmeter to channel 2 again, and enter the measured signal for
channel 2.
Highest possible voltages
Enter - Analog output channel 1 :
3.6
Volt
Enter - Analog output channel 2 :
3.6
Volt
< ESC > aborts
7.
Enter the analog output scaling (Default: 5mV/
°
C, 0V=0
°
C).
Analog fixpoints
Gain
:
5
mV/
°
C
Offset
:
0
°
C
< ESC > aborts