background image

User Manual

5

Model DT-8300EX

Technical Specification

Model

DT-8300EX

Type

Micro-controller based Digital Battery-powered temperature Indicator

Input Signal

Thermocouple sensor or RTD Pt 100 (3 – wire) sensor.

Thermocouple type

J (Fe-Ko element)

K (Cr-Al element) 

Indication

4 digit seven-segment LCD display

Indication Accuracy

Better than ±0.5 °C

Display Height

12.5 mm (1/2”)

Scale Range

-999 to 9999

Calibrated Scale

K-type Thermocouple (-50 to 1000 °C)

J-type Thermocouple (-50 to 750 °C)

RTD Pt-100 (3-wire) (-100 to 500 °C)

Over-Range Indication

Err [indicating “Error”]

Power Supply

Two batteries, type AA, 3VDC

Switch Selections

Power (On / Off)

Temperature units (°C / °F)\

Type of Sensor

Type of Thermocouple.

Cold Junction Compensation

Provided

Response Time

Typically 75 mS

Display Update Time

10 Seconds

Linearization

Provided for both TC types, eight-point

Minimum Resolution

0.1 Degree C/F

SPECIFICATIONS

CONDITION OF ACCEPTABILITY

This equipment must be powered by a Non-Hazardous live source in accordance with 
CAN/CSA-C22.2 No. 61010-1-04 (eg. SELV source), or a Class 2 Source.

Summary of Contents for DT-8300EX

Page 1: ...www ametekusg com User Manual BATTERY OPERATED DIGITAL TEMPERATURE INDICATOR Installation and Operating Instructions for the Model DT 8300EX...

Page 2: ...umidity including condensation and dust The instrument is micro controller based and operates on battery power having provision for two Type AA batteries in series configuration It is specifically des...

Page 3: ...ing the instrument The maximum recommended distance of the Temperature Sensor from the indicating instrument is 15 meters OPERATION Remove the electronics from the outer housing and loosen the two scr...

Page 4: ...lector switch which allows the user to display temperature in either degrees C or degrees F The OFF position of this switch will display degrees F BATTERY OPERATED DIGITAL TEMPERATURE INDICATOR 17348...

Page 5: ...rees for any corrections in non linearity or errors on account of long Cables Sensor faults etc This indicates an Error condition It could mean either a broken sensor or open sensor or abnormally high...

Page 6: ...as a scan Cycle Time of 10 seconds to extend Battery life This means that the LCD display is updated every ten seconds Because of this the keys may appear to respond slowly at times BATTERY OPERATED D...

Page 7: ...ple 50 to 1000 C J type Thermocouple 50 to 750 C RTD Pt 100 3 wire 100 to 500 C Over Range Indication Err indicating Error Power Supply Two batteries type AA 3VDC Switch Selections Power On Off Temper...

Page 8: ...rranty set forth herein does not extend to and shall not apply to Products which have been repaired or altered by other than AMETEK s personnel unless Buyer has properly altered or repaired the produc...