Universal IV™ CM Density Compensated Model
52
Code
Display Message
Description
Actions
ERROR 01
CALCULATION K EFF EXCEEDS
K H2O ERROR
Calculated effective dielectric
constant exceeds water dielectric
constant
1. Cycle power.
2. Hardware Failure - Contact
factory
ERROR 02
CALCULATION EFF CAP
RANGE ERROR
Calculated effective capacitance
is outside of expected limits
1. Calibrate DCM
2. Verify U-IV CM is operating
correctly. Reference U-IV CM
manual
3.Contact factory
ERROR 03
CALCULATION AIR CAP RANGE
ERROR
Calculated air capacitance is
outside of expected limits
1. Verify U-IV CM & Coriolis are
operating correctly.
2. Verify Cut Monitor is for the
correct pipe size.
3.Contact factory
ERROR 04
CALCULATION EFF DENSITY
RANGE ERROR
Calculated effective (mixture)
density is outside of expected
limits
1. Verify U-IV CM & Coriolis are
operating correctly
2.Contact factory
ERROR 05
CALCULATION H2O DENSITY
RANGE ERROR
Calculated water density is
outside of expected limits
1. Verify U-IV CM & Coriolis are
operating correctly
2.Contact factory
ERROR 06
CALCULATION OIL DENSITY
RANGE ERROR
Calculated oil density is outside of
expected limits
1. Verify U-IV CM & Coriolis are
operating correctly
2.Contact factory
ERROR 07
CALCULATION TEMPERATURE
RANGE ERROR
Calculated temperature is outside
of expected limits
1. Verify U-IV CM & Coriolis are
operating correctly
2.Contact factory
ERROR 08
CALCULATION K EFF RANGE
ERROR
Calculated effective dielectric
constant is outside of expected
limits
1. Verify U-IV CM & Coriolis are
operating correctly
2.Contact factory
ERROR 09
CALCULATION K EFF RANGE
ERROR
Calculated water dielectric
constant is outside of expected
limits
1. Verify U-IV CM & Coriolis are
operating correctly
2.Contact factory
ERRPR 10
CALCULATION FLOWRATE
THRESHOLD ERROR
Flow rate has dropped below
threshold. Density value is locked
on last density before threshold
drop
1. If system is not running, no
action required
2. Verify Coriolis operating
properly
3. Consult factory
ERROR 11
FIELD FLASH CKSM ERROR
Field flash checksum failure Lost
customer configuration
Reconfigure DCM
ERROR 12
IM INTERMITTENT COMM LOSS
ERROR
Inter microprocessor intermittent
loss of communication
No Action required
5.3 Error Codes (Continued)
Table 5.2
Summary of Contents for Drexelbrook Universal IV CM
Page 4: ...Contents...
Page 6: ...Page Has No Content...
Page 7: ...Section 1 Introduction...
Page 9: ...Introduction 2 1 3 Model Numbering 1 3 1 System Electronics Probe DCM Module Model UDO System...
Page 11: ...Introduction 4 1 3 Model Numbering Continued 1 3 3 In Line Spool Piece Probe...
Page 13: ...Section 2 Installation...
Page 28: ...Installation 21 2 5 Modbus Validation Continued...
Page 29: ...Universal IV CM Density Compensated Model 22 2 5 Modbus Validation Continued...
Page 30: ...Section 3 Configuration Calibration...
Page 37: ...Universal IV CM Density Compensated Model 30 3 4 Menu Tree...
Page 52: ...Section 4 Specifications...
Page 54: ...Section 5 Normal Maintenance...
Page 61: ...Page Has No Content...
Page 62: ...Section 6 Hazardous Locations...
Page 67: ...Page Has No Content...
Page 68: ...Section 7 Control Drawings...
Page 69: ...Universal IV CM Density Compensated Model 62 7 1 ATEX IECEX Section 7 0 Control Drawings...
Page 70: ...Control Drawings 63 7 1 ATEX IECEX Continued...
Page 71: ...Universal IV CM Density Compensated Model 64 7 1 ATEX IECEX Continued...
Page 72: ...Control Drawings 65 7 1 ATEX IECEX Continued...
Page 73: ...Universal IV CM Density Compensated Model 66 7 1 ATEX IECEX Continued...
Page 74: ...Control Drawings 67 7 1 ATEX IECEX Continued...
Page 75: ...Universal IV CM Density Compensated Model 68 7 1 ATEX IECEX Continued...
Page 76: ...Control Drawings 69 7 1 ATEX IECEX Continued...
Page 77: ...Universal IV CM Density Compensated Model 70 7 2 FM US FMC...
Page 78: ...Control Drawings 71 7 2 FM US FMC Continued...
Page 79: ...Universal IV CM Density Compensated Model 72 7 2 FM US FMC Continued...
Page 80: ...Control Drawings 73 7 2 FM US FMC Continued...
Page 81: ...Universal IV CM Density Compensated Model 74 7 2 FM US FMC Continued...
Page 82: ...Control Drawings 75 7 2 FM US FMC Continued...
Page 83: ...Universal IV CM Density Compensated Model 76 7 2 FM US FMC Continued...
Page 84: ...Control Drawings 77 7 2 FM US FMC Continued...
Page 85: ...Universal IV CM Density Compensated Model 78 7 2 FM US FMC Continued...
Page 86: ...Control Drawings 79 7 2 FM US FMC Continued...
Page 87: ...Universal IV CM Density Compensated Model 80 7 2 FM US FMC Continued...
Page 88: ...Control Drawings 81 7 2 FM US FMC Continued...
Page 89: ...Universal IV CM Density Compensated Model 82 7 2 FM US FMC Continued...
Page 90: ...Control Drawings 83 7 2 FM US FMC Continued...
Page 91: ...Universal IV CM Density Compensated Model 84 7 2 FM US FMC Continued...
Page 92: ...Control Drawings 85 7 2 FM US FMC Continued...
Page 93: ...Universal IV CM Density Compensated Model 86 7 2 FM US FMC Continued...
Page 94: ...Control Drawings 87 7 2 FM US FMC Continued...
Page 95: ...Universal IV CM Density Compensated Model 88 7 2 FM US FMC Continued...
Page 96: ...Control Drawings 89 7 2 FM US FMC Continued...
Page 97: ...Page Has No Content...
Page 98: ...Section 8 Approval Certificates...
Page 100: ...Approval Certificates 93 8 1 FM US Approval Certificate Continued...
Page 101: ...Universal IV CM Density Compensated Model 94 8 1 FM US Approval Certificate Continued...
Page 102: ...Approval Certificates 95 8 1 FM US Approval Certificate Continued...
Page 103: ...Universal IV CM Density Compensated Model 96 8 1 FM US Approval Certificate Continued...
Page 104: ...Approval Certificates 97 8 1 FM US Approval Certificate Continued...
Page 105: ...Universal IV CM Density Compensated Model 98 8 1 FM Canada Approval Certificate...
Page 106: ...Approval Certificates 99 8 1 FM Canada Approval Certificate Continued...
Page 107: ...Universal IV CM Density Compensated Model 100 8 1 FM Canada Approval Certificate Continued...
Page 108: ...Approval Certificates 101 8 1 FM Canada Approval Certificate Continued...
Page 109: ...Universal IV CM Density Compensated Model 102 8 1 FM Canada Approval Certificate Continued...
Page 110: ...Approval Certificates 103 8 1 FM Canada Approval Certificate Continued...
Page 111: ...Universal IV CM Density Compensated Model 104 8 2 IECEx Approval Certificate...
Page 112: ...Approval Certificates 105 8 2 IECEx Approval Certificate Continued...
Page 113: ...Universal IV CM Density Compensated Model 106 8 2 IECEx Approval Certificate Continued...
Page 114: ...Approval Certificates 107 8 2 IECEx Approval Certificate Continued...
Page 115: ...Universal IV CM Density Compensated Model 108 8 2 ATEX Approval Certificate...
Page 116: ...Approval Certificates 109 8 2 ATEX Approval Certificate Continued...
Page 117: ...Universal IV CM Density Compensated Model 110 8 2 ATEX Approval Certificate Continued...
Page 118: ...Approval Certificates 111 8 2 ATEX Approval Certificate Continued...
Page 119: ...Universal IV CM Density Compensated Model 112 8 2 ATEX Approval Certificate Continued...
Page 120: ...Approval Certificates 113 8 2 ATEX Approval Certificate Continued...
Page 121: ...Universal IV CM Density Compensated Model 114 8 2 ATEX Approval Certificate Continued...
Page 122: ...Approval Certificates 115 8 3 CE Mark Declaration of Conformity...
Page 123: ...Page Has No Content...
Page 124: ...Section 9 Appendix...
Page 126: ...Section 10 Terms Conditions...