After selecting “RECORD”: if other parameters had previously been changed, the query
‘ACCEPT YES” is first made as to whether they are to be stored. In this case, confirm by
pressing 

. To record, use the

 ↓ 

key to select one of the following options:

• 

MAX. VALUES: (only maximum values are taken into account when the empty-tank

spectrum is recorded; useful e.g. with “difficult“ agitators).

• 

AVERAGE: (values are averaged; this setting can be used for most applications).

After selecting with the 

↓ 

key, press the 

↵ 

key to select TOTAL or the 

↓ 

key to select

PARTIAL.

• 

When TOTAL is selected, the empty-tank spectrum is recorded over the entire range

(tank height).

• 

If the tank has not been fully drained, the empty-tank spectrum can also be

recorded up to a certain distance, in which case the menu item PARTIAL should be
selected. When this has been selected, a query takes place by way of the 

↵ 

key

concerning the distance value up to which the empty-tank spectrum is to be
recorded. The tank area below the current filling level is then excluded from the
empty-tank recording. It is recommended to maintain a safety distance of 8 to 12 inches
from the actual product distance.

Subsequently press key 

↵ 

to start recording the empty-tank spectrum. The display starts

with “200“ and counts down to “0“. The sign WAIT... is shown in the display. READY
appears after approximately 1 to 3 minutes. Then press key 

↵ 

five times to store the re-

corded empty-tank spectrum, which is taken into account for measurements.

Empty-tank spectrum recording via PC-CAT
Connect the DR22X0 and press in the display mode of PC-CAT the key combination
Ctrl-L. The type of empty-tank spectrum can be selected by one of the following keys:
1: Max.Values 4: Max. Partial A: Break
2: Average 5: Avg. Partial

Tank bottom tracing mode (FTB)

The DR22X0 includes an additional function for measuring reliably low levels in tanks with
flat bottom and poorly reflecting products (low dielectric constant). This tank bottom tracing
system (abbreviated FTB) is activated in the vicinity of the tank bottom (max. 20% level).
Given higher levels, the normal measuring method is used (reflection from the product
surface). If the measurement jumps to the correct level only after filling above a certain level
(approximately 0.3-1.0 m), you can activate the FTB function 

Fct. 3.5.7 

“PARTIAL“. The

relative permittivity 

ε

r of the tank product must be set in 

Fct. 3.5.8

. If it is not known, enter

the figure of 2.0. Since the exact position of the tank bottom must be known for this pro-
cess, it is advisable when using the FTB to determine the tank height automatically with an
empty tank, use 

Fct. 3.5.1

.

Conversion table/Volume table

A table consisting of a maximum of 50 points can be stored in the DR22X0 for non-linear or
linear conversion of the level, e.g. into a volumetric value. This table, however, can only be
programmed with the PC-CAT program (Fct. 3.7.2).

Summary of Contents for DR22X0 Series

Page 1: ...http www drexelbrook com e mail Service Drexelbrook com Leader in Level Measurement For Assistance Call 1 800 553 9092 Outside North America 1 215 674 1234 Installation and Operating Instructions Ser...

Page 2: ...ce or use of material Copyright 2001 AMETEK Drexelbrook The DR2200 level gauge is designed solely for measuring the distance level volume and reflection of liquids solids and particulate materials The...

Page 3: ...Service 800 527 6297 International 215 674 1234 Fax 215 674 2731 E mail deinfo drexelbrook com Web www drexelbrook com An ISO 9001 Certified Company 205 Keith Valley Road Horsham PA 19044 Series DR22X...

Page 4: ...1 2 1 Field Assembly 1 2 2 Mechanical Installation 3 SECTION 3 Electrical Connection 4 SECTION 4 Setting the Parameters 5 SECTION 5 Maintenance error handling 15 SECTION 6 Safety Information 17 SECTI...

Page 5: ...s are supplied in fully assembled condition However if a de vice should be delivered in parts or parts are subsequently replaced use the following assembly information 2 1 Field assembly 1 For any nec...

Page 6: ...TFE used in all versions with the product See also section 8 Type code Mounting on the tank nozzle a Devices with horn antenna The antenna should project out of the nozzle If necessary use an antenna...

Page 7: ...DR22X0 and gasket slightly tighten nuts on stud bolts by hand Tighten down stud bolt nuts firmly The tightening torque is dependent upon the strength properties of the stud bolts and the pressure rati...

Page 8: ...on 6 Supply voltage at the terminals 1 2 The 4 20 mA supply must be able to provide the following voltage U at the terminals of the DR22X0 dependent on the current I Please consider also the line resi...

Page 9: ...he isolation amplifier over a load between 120 and 350 to the Smart adapter and connect it with a serial port of the PC The isolation amplifier must be HART compatible Local display optional Adapter 1...

Page 10: ...formation about the status of measurement and are no error displays 1 No current measured value The device is searching for a new value If the search for a plausible level fails for a certain time SIG...

Page 11: ...value shall not be stored abort by Or accept this 4mA scaling by pressing SURE YES and then 8 Fill the tank to the 100 mark Use the same procedure for the 100 point 20 mA only now by pressing the top...

Page 12: ...ut of selected value to the current output With safety query 2 4 FIRMWARE 2 4 1 MASTER Display Display of master firmware version 3 0 INSTALL 3 1 BASIS PARAM 3 1 1 TANKHEIGHT Select unit m cm mm inch...

Page 13: ...le 3 2 5 ERROR MSG Select NO YES Select whether error messages to be shown in display 3 3 SIGNAL OUT 3 3 1 FUNCTION I Select OFF LEVEL DISTANCE CONVERSION SW OUTP Select function of the current output...

Page 14: ...e 1 10 100 s Enter time constant for measured value filtering 3 5 4 TRACING VEL Value 0 01 0 50 10 00 m Min Enter the maximum rate of change in level that can occur in operation 3 5 5 MULT REFL Select...

Page 15: ...KH Before confirming check carefully that the proposed tank height is within range Block distance The block distance function Fct 3 1 2 defines a zone below the top reference point in which measuremen...

Page 16: ...etely empty and all moving parts e g agitators switched on If major interference through internals is not expected recording of the empty tank spectrum is not necessary since the factory has already c...

Page 17: ...ars after approximately 1 to 3 minutes Then press key five times to store the re corded empty tank spectrum which is taken into account for measurements Empty tank spectrum recording via PC CAT Connec...

Page 18: ...ock distance 0 60 m 0 6000 m Confirm block distance and move to current output configuration Fct 3 3 SIGNAL OUT Move to lower range value Fct 3 3 3 SCALE 4 mA Display default value 00 000 m Enter lowe...

Page 19: ...parameters are correct for your application If not reset 6 Record the empty spectrum see Section 4 Important Ensure that the screw threads on the terminal and electronic compartments are well greased...

Page 20: ...terial Safety Data Sheets MSDS listing the hazardous materials that the sensing element has been exposed to must accompany any repair It is your responsibility to fully disclose all chemicals and deco...

Page 21: ...ismantling or making electrical connections in a hazardous area Temperature rating of connecting cables The temperature rating of connecting cables is dependent on the maximum temperature of the flang...

Page 22: ...nce reflections and multiple reflections Hazardous locations DR22X0 Ex Zone 0 1 2 IIC IIB T1 toT6 Intrinsically safe Class 1 Div 1 and 2 Ambient temperature at signal converter 4 to 131 F 20 to 55 C f...

Page 23: ...elloy C4 or B2 titanium tantalum Other materials available on request Gaskets FFKM Kalrez 4079 or Parofluor V 8545 75 Kalrez 2035 Viton FPM FEP coated PTFE is also in contact with the product Wave Sti...

Page 24: ...ll B Wave Guide 59 inches 3 Horn Type 3 6 inch standard C Wave Guide 78 inches 4 Horn Type 4 8 inch Power supply 4 2 wire loop powered Output 2 4 to 20 mA passive HART I S Display 0 Without 1 With Cab...

Page 25: ...i Clamp M 3 Tri Clamp N 4 Tri Clamp 0 Without 1 PTFE Rod antenna 15 1 max 232psi 2 PTFE Rod antenna 19 7 max 232psi 3 PTFE Rod antenna 23 6 max 232psi 4 PTFE Rod antenna 27 6 max 232psi 5 PTFE Rod ant...

Page 26: ...3 1 4 Antenna extension 3 1 5 Distance piece 3 1 6 Stillwell diameter 3 1 7 Reference offset 3 1 8 Tank bottom 3 3 1 Current output function offset 3 3 2 Current output range error 3 3 3 Min current s...

Page 27: ...800 553 9092 24 Hour Service 800 527 6297 International 215 674 1234 Fax 215 674 2731 E mail deinfo drexelbrook com Web www drexelbrook com An ISO 9001 Certified Company 205 Keith Valley Road Horsham...

Reviews: