SECTION 4
–
MAINTENANCE AND SERVICING 4-1
Section 4 - Maintenance and Servicing
The 7360V Curing Chamber was designed to require little maintenance and servicing. The
following guidelines are recommended to help insure years of trouble free service. In the
event a problem occurs, a troubleshooting guide is listed following the maintenance
guidelines.
Maintenance
Make certain that the plug and cylinder threads are clean and well lubricated with Moly
grease or other high temperature lubricant.
Check the internal plug O-ring periodically for cuts, tears, or deterioration. Replace if any of
these are evident.
Note: You must install a new O-ring after any test in which the temperature exceeds
350°F/177°C.
Clean all cement out of the bottom of the cylinder.
Troubleshooting
The table on the following page is intended to help identify the source of any malfunctions
that may occur with the 7360V Vertical Curing Chamber.
Summary of Contents for CHANDLER ENGINEERING 7360V
Page 8: ......
Page 14: ......
Page 22: ......
Page 24: ...CHANDLER ENGINEERING...
Page 25: ......