SECTION 2
–
OPERATING INSTRUCTIONS 2-1
Section 2
–
Operating Instructions
Operation
The control enclosure includes four program buttons (P1-P4) located just below the display.
All program buttons are preset at 4,000 RPM for 15 seconds followed by 12,000 RPM for 35
seconds. These buttons may be programmed for up to 8 speed / time combinations (1,000 to
20,000 RPM and 0 to 999 seconds). Refer to the
Maintenance
section of this manual for
additional information on how to reprogram the functionality of these buttons.
Alternatively, starting the mixer via the Start/Stop button allows the speed to be manually
controlled via the Speed +/- buttons for continuous operation of the motor (up to 9999
seconds).
Front Panel Description
1.
LCD Display
–
Displays current speed, mixing time and allows setup, programming,
etc.
2.
Program Buttons
–
Allows the mixer to operate at preset speeds for preset time
intervals. Factory default settings for all Program Buttons are as follows:
a.
4,000 RPM for 15 seconds
b.
12,000 RPM for 35 seconds
3.
Power On/Off (Standby) Button
–
Turns the power on/off.
4.
Manual Speed Control Buttons
–
Controls the mixer speed. If idle, the Start/Stop
button must be pressed to run at the prescribed manual speed.
5.
Time +/- Buttons
–
Used for configuring the mixer and setting program parameters.
6.
Start/Stop Button
–
Stops the mixer if it is currently running. Starts the mixer in
manual speed control mode if it is currently stopped.
Summary of Contents for Chandler Engineering 3260
Page 4: ...T 2 TABLE OF CONTENTS This page is intentionally left blank...
Page 8: ...P 4 PREFACE This page is intentionally left blank...
Page 10: ...1 2 SECTION 1 INSTALLATION This page is intentionally left blank...
Page 18: ...4 2 SECTION 4 REPLACEMENT PARTS 3260 REPL_PARTS REV B This page is intentionally left blank...
Page 20: ...5 2 SECTION 5 DRAWINGS AND SCHEMATICS 3260 DWGS REV C This page is intentionally left blank...