Kalibrationsprozedur gem. IEC 61000-4-12 Ed.3
6.5.1. Ringwave Impulsformdefinition gem. IEC 6100-4-12 Ed.3
Parameter
voltage rise time
T
1
:
0,5
µ
s ± 30 % (open-circuit)
current rise time
T
1
:
≤
1
µ
s (short-circuit)
voltage osc frequency: 100 kHz ± 10 %;
decaying Voltage only 0,4 < Ratio of
Pk
2
to
Pk
1
< 1,1
0,4 < Ratio of
Pk
3
to
Pk
2
< 0,8
0,4 < Ratio of
Pk
4
to
Pk
3
< 0,8
output impedance:
12
Ω
and 30
Ω
± 20 % (switchable)
open circuit voltage
Pk
1:
250 V to 4 kV (± 10 %)
short-circuit current
Pk
1:
333 A ± 10 % for 12
Ω
gen. setting
133 A ± 10 % for 30
Ω
gen. setting
Ring Wave pulse, Open circuit
Parameter
CD
Kopplung
Einheit
Nominalwert
Toleranz
Gerätemodell
- - -
- - -
- - -
N1, N1.1
N2, N2.1
- - -
Pulslevel max.
CAP
Lx-PE, Lx-Lx
[V]
4000
6000
10%
Standard Levels
CAP
Lx-PE, Lx-Lx
[V]
250
500
1000
2000
10%
Rise time 1. Peak
CAP
Lx-PE, Lx-Lx
[
s]
0.5
30%
Oszillation Frequenz
CAP
Lx-PE, Lx-Lx
[kHz]
100
10%
Abklingen vom Peak 2
CAP
Lx-PE, Lx-Lx
[%]
40% - 110%
of Peak 1
- - -
Abklingen vom Peak 3
CAP
Lx-PE, Lx-Lx
[%]
40% - 80%
of Peak 2
- - -
Abklingen vom Peak 4
CAP
Lx-PE, Lx-Lx
[%]
40% - 80%
of Peak 3
- - -
Restspannung
CAP
Lx-PE
[V]
15% of Vpeak max or twice the rated
peak voltage of the coupling / decoupling
network, whichever is higher
- - -
Ring Wave pulse Short circuit
Parameter
CD
Kopplung
Einheit
Nominalwert
Toleranz
Gerätemodell
- - -
- - -
- - -
N1, N1.1
N2, N2.1
- - -
Pulslevel max.
CAP
Lx-PE, Lx-Lx
[V]
4000
6000
- - -
Strom I bei 12
CAP
Lx-PE, Lx-Lx
[A]
333.3
500.0
10%
Strom I bei 30
CAP
Lx-PE, Lx-Lx
[A]
133.3
200.0
10%
Standard Level
CAP
Lx-PE, Lx-Lx
[V]
250
500
1000
2000
10%
Strom I bei 12
CAP
Lx-PE, Lx-Lx
[A]
20.8
41.7
83.3
166.7
10%
Strom I bei 30
CAP
Lx-PE, Lx-Lx
[A]
8.3
16.7
33.3
66.7
10%
Rise time 1. Peak
CAP
Lx-PE, Lx-Lx
[
s]
≤1
- - -
Oszillation Frequenz
CAP
Lx-PE, Lx-Lx
[kHz]
100
10%
Ring Wave Impedanz
Coupling
CD
Impedanz
Einheit
Nominalwert
Toleranz
Leitung - PE,
Leitung - Leitung
CAP
12
[
]
12
20%
Leitung - PE,
Leitung - Leitung
CAP
30
[
]
30
20%