Installation test
6
ÄKTAexplorer Installation Guide 18-1139-59 Edition AC
25
10 When the test is finished, the printer automatically prints the
chromatogram and the test result.
6.5
Evaluating the installation test results
6.5.1
Automatic evaluation
The system automatically prints the test result when the test is finished.
The print-out consists of a chromatogram and an evaluation of the test
result.
•
If the gradient test result is OK, the print-out says
“Gradient linearity accepted”.
•
If the step response test result is OK, the print-out says
“Step response accepted”.
•
If the UV response test result is OK, the print-out says
“UV response accepted”.
If any of the evaluated values fall outside the specified range, go to step
6.6 Correcting faulty evaluation results.
Gradient result
Step response result
100%
95%
70%
30%
0%
5%