11
98-0156 Rev G
Note: The maximum number of dialysis machines that the unit can
supply depends on the volume of hot water required per machine and
the rate at which they draw water. So long as the total volume of water
required does not exceed the maximum available for draw off and the
maximum draw rate does not exceed the pump performance. See
pump limits above.
Environment
Parameter
Limits
Room storage and operating
temperature range
(41-104°F)
Relative humidity
30 to 80%
Maximum altitude
6562 ft (2000m)
Transport and storage
temperature
23°F to 203°F (Dry unit)
RFI/EMI radiation
The EMC environment must be within the
limits to which the unit has been tested, see
Section
below.
Care must be taken not to place near the unit
any source of RFI/EMI, which is liable to cause
electromagnetic disturbance. If the unit is
affected by such disturbance, the source must
be suppressed or moved.
Classification and Standards Applied
Class 1 equipment
Electrical
Insulation
Category II
Pollution
Degree II
Type IIB applied part
EMC
UL 61010-1, Issue 07/12/2004, Ed:2,
Rev:10/28/2008 Standard for Electrical
Equipment for Measurement, Control,
and Laboratory Use Part 1: General
Requirements and CSA C22.2# 61010-1,
Issue: 07/12/2004, Ed:2 (R2009) Safety
Requirements for Electrical Equipment
for Measurement, Control, and
Laboratory Use Part 1: General
Requirements, with general instruction
No. 1:2008/10/28 – R2009.
Summary of Contents for Heatsan 00HS208
Page 51: ...50 98 0156 Rev G Heatsan Process Flow Diagram ...
Page 52: ...51 98 0156 Rev G Wiring Diagrams ...
Page 53: ...52 98 0156 Rev G ...
Page 54: ...53 98 0156 Rev G ...
Page 55: ...54 98 0156 Rev G ...
Page 56: ...55 98 0156 Rev G ...
Page 57: ...56 98 0156 Rev G ...
Page 58: ...57 98 0156 Rev G Installation Layout Water Connection Detail ...
Page 62: ...61 98 0156 Rev G Tubing with Heatsan and MediQA Hidden ...
Page 64: ...63 98 0156 Rev G RO Divert Valve and Heatsan Divert Valve 1 Heatsan Divert Valve 2 ...
Page 65: ...64 98 0156 Rev G Heatsan Discharge Solenoid SV3A NC Heatsan Fill Solenoid SV3B NO ...
Page 70: ...69 98 0156 Rev G ...