7. 

Disconnect the pump outlet hose from the CIP drum and direct the hose to a 
drain. Continue running to drain until the CIP drum is empty (some solution 
may need to be manually dumped out of the CIP drum). 

CAUTION: Neutralization of the spent cleaning solution may be required by 

local regulations. Verify the requirements of local regulations 
prior to emptying the cleaning solution into the drain.

8. 

After removing the spent chemical from the CIP tank, and with the CIP 
pump discharge directed to drain, turn on the MRO and allow fresh water to 
flow into and through the RO and CIP system for 10 minutes or until the pH 
of the water at the pump discharge is neutral (between 6 and 8). 

9. 

Disconnect the pump outlet hose from the CIP and direct it to a drain. 
Continue running to drain until the CIP drum is empty (some solution may 
need to be manually dumped out of the CIP drum). 

FLUSH AND RETURN TO SERVICE 

1.  Disconnect the RO+ Product Water For Dialysis hose, Reject To Drain hose, 

and Incoming Tap Water hose from the CIP system. 

2.  Run the RO+ Product Water Hose and the Reject Water To Drain Hose to a 

drain. 

3.  Reconnect the carbon tank (if applicable) to the RO+. 

4.  Install a new micron filter in the prefilter sump on the RO+. 

5.  Place the CIP Switch located on the back of the RO+ controller in the OFF 

position (if applicable).  Reference the RO+ Operation and Maintenance 
Manual for detailed instructions regarding the Clean-In-Place Switch. 

6.  Connect the RO+ Incoming Tap Water hose to a water supply and turn on the 

incoming water supply. 

7.  Turn on the RO+ and place the FLUSH VALVE in the FLUSH position.  Allow 

the system to run at this setting for 10 minutes. 

8.  Place the FLUSH VALVE in the IN SERVICE or OPERATION position and 

continue running to drain until the conductivity returns to normal (below 
setpoint and not in alarm). 

9.  The RO+ may now be connected to the dialysis machine and returned to 

service. 

WARNING:  DO NOT connect the RO+ to the dialysis machine and/or place 

the RO+ in service until the Product Water conductivity 
remains below the setpoint and is not in alarm!  Serious injury 
or illness to the patient may result! 

Summary of Contents for CIP1

Page 1: ...Part Number 98 0007 Revision E CLEAN IN PLACE OPERATION MAINTENANCE INSTRUCTIONS MODEL CIP1 Manufactured With Pride In The USA www ameriwater com 800 535 5585 AmeriWater 3345 Stop 8 Rd Dayton OH 45414...

Page 2: ......

Page 3: ...TABLE OF CONTENTS OBJECTIVE 1 CHEMICALS REQUIRED 1 INSTALLATION 2 HIGH pH CLEANING 3 LOW pH CLEANING 5 FLUSH AND RETURN TO SERVICE 6 SPARE PARTS LIST 7...

Page 4: ...oulants from polyamide thin film composite membranes 1 lb 372 ml of AmeriClean B powder is required per membrane cleaning AmeriClean B effectively removes the following foulants Silt Microbiological C...

Page 5: ...E 1 5 Connect the RO Incoming Tap Water Hose to the male garden hose thread fitting on the CIP Pump Outlet FIGURE 2 6 Remove the carbon tank if applicable from the RO and install the provided jumper f...

Page 6: ...s dissolved before continuing WARNING Wear appropriate eye and skin protection when handling the membrane cleaning chemicals The cleaners are high pH and low pH chemicals that may cause severe chemica...

Page 7: ...direct it to a drain Continue running to drain until the CIP drum is empty some solution may need to be manually dumped out of the CIP drum WARNING Never allow AmeriClean A and AmeriClean B to mix Mix...

Page 8: ...hemicals and may cause severe chemical burns Read the MSDS for further information 4 Connect the CIP pump power cord to a 120 VAC receptacle GFI recommended 5 Place the Flush Valve in FLUSH then turn...

Page 9: ...ect To Drain hose and Incoming Tap Water hose from the CIP system 2 Run the RO Product Water Hose and the Reject Water To Drain Hose to a drain 3 Reconnect the carbon tank if applicable to the RO 4 In...

Page 10: ...14570110 WASHER RUBBER GH 14570111 FITTING 3 4 MGHT x 1 2 MPT 046520423 HOSE 1 2 MEDICAL SYTLE 5000 12677125 CLAMP HOSE 1 2 SST 15650050 PUMP AQUATEC 120VAC 3 5GPM 50PSI 80 0026 ADPT AQUAPUMP TO 5MPT...

Page 11: ...8...

Reviews: