LOW pH CLEANING 

AmeriClean A

 is used second for the removal of inorganic scale and metals. 

1. 

Add 10 gallons (~2 inches 
below the top hose connection) 
of clean water to the CIP drum 
(FIGURE 3).  RO+ product 
water or deionized water is 
recommended if available. 
Water with a minimum 
temperature at or near 86° F 
will provide maximum 
effectiveness for the high pH 
cleaner. 

2. 

Slowly add 1 lb (1 full bag) of 

AmeriClean A

 to the water in the cleaning 

drum. 

3. 

Agitate the solution until the cleaner is completely dissolved. 

It is important 

to make sure all of the powder is dissolved before continuing.

WARNING:  Wear appropriate eye and skin protection when handling the 

membrane cleaning chemicals.  The cleaners are high and low 
pH chemicals and may cause severe chemical burns.  Read the 
MSDS for further information. 

4. 

Connect the CIP pump power cord to a 120 VAC receptacle (GFI 
recommended). 

5. 

Place the Flush Valve in “FLUSH” then turn on the RO+ and verify that the 
cleaning solution is recirculating and flowing from the drain hose inside the 
CIP drum. 

6. 

Allow the cleaning solution to recirculate for a minimum of 50 minutes. 

a. Depending on the nature of the fouling, a soak period may be necessary. 

If the membranes are not heavily fouled, proceed to step 7. 

b. In cases of heavy fouling, turn off the RO and CIP pump and allow the 

system to soak for a minimum of 30 minutes.*   

*  Contact AmeriWater Technical Support or your local AmeriWater dealer for custom cleaning procedures if 

necessary. 

c.  After the soak time, turn on the CIP pump and RO, and allow the cleaning 

solution to recirculate again for 30 minutes.

FILL TO 2” BELOW 

TOP HOSE 

CONNECTIONS 

FIGURE 3 

Summary of Contents for CIP1

Page 1: ...Part Number 98 0007 Revision E CLEAN IN PLACE OPERATION MAINTENANCE INSTRUCTIONS MODEL CIP1 Manufactured With Pride In The USA www ameriwater com 800 535 5585 AmeriWater 3345 Stop 8 Rd Dayton OH 45414...

Page 2: ......

Page 3: ...TABLE OF CONTENTS OBJECTIVE 1 CHEMICALS REQUIRED 1 INSTALLATION 2 HIGH pH CLEANING 3 LOW pH CLEANING 5 FLUSH AND RETURN TO SERVICE 6 SPARE PARTS LIST 7...

Page 4: ...oulants from polyamide thin film composite membranes 1 lb 372 ml of AmeriClean B powder is required per membrane cleaning AmeriClean B effectively removes the following foulants Silt Microbiological C...

Page 5: ...E 1 5 Connect the RO Incoming Tap Water Hose to the male garden hose thread fitting on the CIP Pump Outlet FIGURE 2 6 Remove the carbon tank if applicable from the RO and install the provided jumper f...

Page 6: ...s dissolved before continuing WARNING Wear appropriate eye and skin protection when handling the membrane cleaning chemicals The cleaners are high pH and low pH chemicals that may cause severe chemica...

Page 7: ...direct it to a drain Continue running to drain until the CIP drum is empty some solution may need to be manually dumped out of the CIP drum WARNING Never allow AmeriClean A and AmeriClean B to mix Mix...

Page 8: ...hemicals and may cause severe chemical burns Read the MSDS for further information 4 Connect the CIP pump power cord to a 120 VAC receptacle GFI recommended 5 Place the Flush Valve in FLUSH then turn...

Page 9: ...ect To Drain hose and Incoming Tap Water hose from the CIP system 2 Run the RO Product Water Hose and the Reject Water To Drain Hose to a drain 3 Reconnect the carbon tank if applicable to the RO 4 In...

Page 10: ...14570110 WASHER RUBBER GH 14570111 FITTING 3 4 MGHT x 1 2 MPT 046520423 HOSE 1 2 MEDICAL SYTLE 5000 12677125 CLAMP HOSE 1 2 SST 15650050 PUMP AQUATEC 120VAC 3 5GPM 50PSI 80 0026 ADPT AQUAPUMP TO 5MPT...

Page 11: ...8...

Reviews: