background image

controller, respectively. Detailed information regarding their adjustment is given
later in the ADJUSTING THE CONTROLLER section.

LOWER RIGHT (Display Mode)

The position of the SLIDE SWITCH (Fig. F) determines what the LED

display will show. At the center position, the display indicates the temperature as
measured by the probe tip, and the red TEMP LED (Fig. C) at the lower left of the
instrument will be illuminated. Sliding the switch up or down will cause the
display to show the high and low controller setpoints, respectively.

UPPER RIGHT (Controller Status LED’s)

There are a total of 4 LED’s here. The upper row (Fig. G) shows the status

of the 110 VAC outlets on the back of the controller beneath the LCD pH display.
The lower LED lights (Fig. H) are on when the display mode switch is in the HIGH or
LOW position.

110 VAC OUTLETS

There are two outlets located above the LCD display. Outlet #1 (on the

left) will be energized when the temperature is above the High setpoint, and Outlet
#2 (on the right) will be energized when the temperature is below the Low
setpoint. A common application for the controller will be to control a chiller OR a
heater, or a chiller AND a heater. Since the addition of chiller lowers the
temperature you must attach the plug of the chiller to Outlet #1. If a heater is
attached to the controller, it should be plugged into the LOW control Outlet # 2.

A temperature electrode is supplied with the controller.  The probe cable

is ozone resistant silicone. Attach the temperature probe to the BNC connector on
the controller. Replacement temperature probes are available.

IV. PROPER PLACEMENT OF THE TEMPERATURE PROBE

When the controller is operational, it is critical that the probe tip is

immersed in the system at all times. If the water level falls below the sensing tip
of the electrode, the temperature probe will not read properly. If the erroneously
measured temperature is higher than the controller setpoint, the chiller connected

to that outlet will remain permanently ON, irrespective of the actual system
temperature.

Take some time to determine how much the fluid level around the probe

will fluctuate. Adjust the position of the probe accordingly.

Be sure to check the probe position occasionally. Complacency usually

sets in when you feel that “Everything is running fine”.

V. ADJUSTING THE CONTROLLER SETPOINTS
The PINPOINT Temperature Controller is capable of controlling temperature within
the range of 60.0° F to 102.0° F. After the selection of the High and Low
temperature setpoints, the Range screw will create a differential or cushion of
±0.5° F to ±10.0° F above and below each setpoint.

Two adjustment screws on the front panel determine the controller

Temperature setpoints. The right adjustment screw (Center Value Fig. E) should be
adjusted to set the optimal temperature. The left adjustment screw (Range Fig. D)
is used to create the differential both above and below your optimal temperature
setpoint.

Turn the Range screw toward minimum to create the most stable

temperature environment and increase this setting toward maximum to create a
greater cushion around the optimal temperature setting.

Adjusting the Controller Setpoints
When using the PINPOINT Temperature Controller with a chiller put the switch to
the High position. Turn the Range screw toward the Minimum setting as much as
possible for maximum temperature stability. Now adjust the Center Value screw to
select the optimal temperature of your system. Attach the chiller to outlet #1.
Place the selector switch in the middle position to read temperature.
When using the PINPOINT Temperature Controller with a heating device put the
switch in the Low position. Turn the Range screw toward the Minimum setting as
much as possible for maximum temperature stability. Attach the heater to Outlet
#2. Place the selector switch in the middle position to read temperature.

Summary of Contents for PINPOINT

Page 1: ...172 Ridgefield CT 06877 U S A Phone FAX 914 763 5367 www americanmarineusa com PINPOINT Temperature Controller Users Manual American Marine Inc 54 Danbury Road Suite 172 Ridgefield CT 06877 U S A FAX Phone 914 763 5367 www americanmarineusa com ...

Page 2: ......

Page 3: ...m does not mean that you no longer have to pay attention to it You must still monitor the system so that you can note when a problem exists and correct it in a timely fashion II GENERAL SPECIFICATIONS Temperature Measurement Range 32 0 F 212 0 F Temperature Control Range 60 0 F 102 0 F Differential Minimum 0 5 F of a selected temperature Differential Maximum 10 0 F of a selected temperature 3 1 2 ...

Page 4: ...ll not read properly If the erroneously measured temperature is higher than the controller setpoint the chiller connected to that outlet will remain permanently ON irrespective of the actual system temperature Take some time to determine how much the fluid level around the probe will fluctuate Adjust the position of the probe accordingly Be sure to check the probe position occasionally Complacency...

Page 5: ...nc will not be liable for any damage or wear to products or livestock caused by abnormal operating conditions water damage abuse misuse unauthorized alteration or repair or if the product was not installed in accordance with the printed operating instructions Any defective product must be sent freight prepaid with appropriate documentation supporting the warranty claim Replacement or repair will b...

Reviews: