Rev. 5
55
Operation
Setup Menu : Load
performed — all operations will be performed and displayed in
terms of amperes. Values from 0.001 to 999.99999 are acceptable
for coil constant. The default value is 1.00000 kG/A (or 0.10000 T/A)
unless preset by AMI to match a specific superconducting magnet.
If the coil constant is not explicitly stated within a superconducting
magnet’s specifications, the value can be obtained by dividing the
rated field by the rated current. Note that 1 T = 10 kG.
3.10.2.3
Magnet Current Rating
Caution
The
Magnet Current Rating is normally set to match a specific
superconducting magnet. The setting should not be changed unless
a different magnet is to be used; always refer to the magnet
specification before changing the Magnet Current Rating.
A magnet operates within the capabilities of the associated power
supply. Since the supply must be selected from the available
standard ratings, the current limit of the supply, which is not user-
adjustable, is almost always higher than the Magnet Current
Rating.
The Magnet Current Rating is normally preset by AMI to match a
specified superconducting magnet.
If AMI is not supplying the
+0.00 A
P
Magnet Current Rating (A)
+0.00 Vm
±100.000
Figure 3-10.
Typical Power Supply Self-Limits
Summary of Contents for 05100PS-430-601
Page 2: ......
Page 10: ...x Rev 5 List of Figures ...
Page 12: ...xii Rev 5 List of Tables ...
Page 18: ...xviii Rev 5 Foreword Safety Summary ...
Page 30: ...12 Rev 5 Introduction Operating Characteristics ...
Page 42: ...24 Rev 3 Installation Power Up Procedure ...
Page 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Page 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Page 168: ...150 Rev 5 Service Return Authorization ...
Page 226: ...208 Rev 5 Index ...