106
Rev. 5
Remote Interface Reference
SCPI Status System
messages in the output buffers will clear the appropriate “Message Available”
bit. The bit definitions for the Status Byte register are defined in Table 4-1.
Bit 2 of the Status Byte register, indicating a quench condition, remains set
until the quench condition is cleared via the front panel or by remote
command. Bits 3 and 4 remain set until all data has been read from the
respective output buffer.
The Status Byte condition register is cleared when:
•
A
*CLS
command is executed.
•
The Standard Event register is read (only bit 5 of the Status Byte
register is cleared).
•
The indicated condition no longer exists.
The Status Byte enable register is cleared when:
•
The
*SRE 0
command is executed.
•
The power is turned off and then back on, and the Model 430
Programmer was configured for
*PSC 1
(power-on status clear).
The enable register setting is persistent if the Model 430
Programmer is configured for
*PSC 0
(no status clear on power-
on).
Table 4-1.
Bit Definitions for the Status Byte Register
Bit Number
Decimal
Value
Definition
0
Not Used
1
Always “0”.
1
Not Used
2
Always “0”.
2 Quench Condition
4
The Model 430 has detected a
quench.
3 Serial Message
Available
8
The serial output buffer contains
unread data.
4
Not Used
16
Always “0”.
5 Standard Event
32
One or more enabled bits are set in
the Standard Event register.
6 Status Byte
Summary
64
One or more enabled bits are set in
the Status Byte register.
7
Not Used
128
Always “0”.
Summary of Contents for 05100PS-430-601
Page 2: ......
Page 10: ...x Rev 5 List of Figures ...
Page 12: ...xii Rev 5 List of Tables ...
Page 18: ...xviii Rev 5 Foreword Safety Summary ...
Page 30: ...12 Rev 5 Introduction Operating Characteristics ...
Page 42: ...24 Rev 3 Installation Power Up Procedure ...
Page 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Page 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Page 168: ...150 Rev 5 Service Return Authorization ...
Page 226: ...208 Rev 5 Index ...