 11 

 

AV ACQUISITION 4001 / ©AMC VIBRO Sp. z o.o. 

 

of samples, which the AVA module should collect after the measurement is triggered. The module will 
stop working after collecting the defined number of samples. 

The  situation  is  different  if  the  Continuous  option  is  selected,  as  then  the  module  will  feed  data 
continuously to the computer. 

When clicking on the Trigger tab, make sure that the Trigger Enable field is marked. 

4.

 

This area features the Channel name field where you can enter the name of the channel selected. In 
the Analog type field the Voltage option should selected, while in the Units field the selection should 
be Volt. 

5.

 

In  this  field  it  is  possible  to  set  gains  and  so  the  measurement  range.  The  AVA  4001  module  is 
equipped  with  input  amplifiers  configured  by  software,  individually  for  each  channel.  The 
measurement range of each channel may be set to ±10V, ±5V, ±2V or ±1V. Limit values should be 
entered in the Lower limit and Upper limit fields as required. The numbers should be equal in the 
absolute value. When different values are entered, for example -1V and 2V, the program will signal 
an error and will suggest that the greater of them, for example -2V and 2V, is used. 

6.

 

The Current Channel Value field presents values of measurement in progress. 

Once the above parameters are set as required, the AVA module is ready for operation in DASYLAb. A 
description of the use of the program itself is beyond the scope of this manual, whose only purpose is 
to show how to prepare the module for signal processing and  measurement data transmission to a 
computer. 

 

10.

 

Recycling 

10.1.

 

Hazardous Materials  

AVA  4001  system  does  not  use  any  hazardous  materials  outlined  by  RoHS.  These 
regulations  confirm  that  lead,  mercury,  cadmium,  hexavalent  chromium, 
polybrominated  biphenyls,  polybrominated  diphenyl  ether,  or  other  battery  related 
materials are limited to no more than trace amounts.  
 
 

10.2.

 

Recycling Facilities  

When decommissioning out of use devices, minimize the impact of the waste created. Refer to local 
waste removal administration for current information on proper material collection and recycling.  

 

 

 

Summary of Contents for AV ACQUISITION 4001

Page 1: ...MC VIBRO Sp z amc VIBRO Sp z o o ul Pilotow 2e 31 462 Krakow Poland T 48 12 362 97 60 S 48 12 362 97 63 info amcvibro pl KRS No 0000618618 REGON No 364497010 VAT No PL6772403385 www amcvibro pl 4 CHAN...

Page 2: ...3 3 Technical data 4 4 Measurement accuracy 5 5 Description of connections and status lights 7 6 Connection with a computer using Windows 7 7 7 InstaCal 8 8 DASYLab 16 configuration 8 9 Using the AVA...

Page 3: ...to 100 kS s The provided drivers allow operation with DASYLab LabVIEWTM and Matlab Module characteristics operation with four IEPE standard vibration sensors sensor circuit functional check warnings...

Page 4: ...channels is max 400 kS s Measurement range 10V 5V 2V i 1V selected by software Crosstalk 80 dB within 0 25 kHz band Maximum absolute error range 10V 5 66 mV range 5V 2 98 mV range 2V 1 31 mV Range 1V...

Page 5: ...bly small and may be overlooked The device is equipped with a 16 bit ADC converter which transforms analog signals into values ranging from 0 to 65525 The following is an ideal characteristic for a 10...

Page 6: ...elated to the difference between the slope of the real characteristics and the ideal characteristics and is expressed as a percentage For the 10 V range the gain error is 0 04 Adding the offset and th...

Page 7: ...es not require the installation of any drivers in the operating system The operating system detects a new device automatically when connected and displays it in the hardware manager as a USB HID inter...

Page 8: ...AVA module is connected to the computer A window will appear with a list of measurement cards which have been detected Mark the item named USB 1608FS click OK and close InstaCal The above procedure sh...

Page 9: ...w This window needs to be discussed in detail as it gives the user the option to configure the AVA module parameters for operation in DASYLab The following areas can be distinguished 1 You can enter y...

Page 10: ...this window in the Scan rate field it is possible enter required sampling frequency select the proper unit and in the Block size field it is possible to select proper size of data buffer Attention The...

Page 11: ...lds as required The numbers should be equal in the absolute value When different values are entered for example 1V and 2V the program will signal an error and will suggest that the greater of them for...

Reviews: