background image

2.

3.

Connecting the system

Preparation for use

Turn the page

Ambu

®

 aScope

 3 and Ambu

®

 aScope

 3 Slim

- Quick Guide for intubation and single lung ventilation

The aScope 3 and aScope 3 Slim are the perfect choice  
for all types of intubation as they are ready to use  
whenever you need them. The instantly accessible  
single-use and ready-to-go concept simplify flexible 
intubation procedures and improve patient safety.

aScope 3 and aScope 3 Slim are single-use sterile  
flexible videoscopes with a 2.2 mm or 1.2 mm working  
channel respectively.

For intubation the aScope 3 can be used with ET tubes size 
6 and up and the aScope 3 Slim with ET tubes size 5 and up.

For single lung ventilation aScope 3 Slim  can be passed 
through DLT’s of FR 37 or greater or BB of sizes 5-9.

  Ensure that the selected ET tube or  DLT/BB is compatible with 

the aScope 3 or aScope 3 Slim. 

  Lubricate the insertion cord with a medical grade lubricant to 

ensure the lowest possible friction. Do not get lubrication or any 
other contaminant on the tip of the aScope 3/aScope 3 Slim.

  Attach the suction tube to the suction connector and ensure 

that the suction works when pressing the suction button

  Fluids can be instilled through the working channel by inserting a 

syringe into the working channel port at the top of the aScope 3. 
When using a Luer Lock syringe, use the included introducer.

  If you are using a preformed ET tube unbend it before inserting 

the cord of the aScope 3 through it.

  Antifogging agent is not required with this device.

  Confirm that the bending section functions smoothly and 

correctly by carefully sliding the bending control lever forward 
and backward in each direction. 

  Confirm that the bending section returns to a straight position.

  Connect the aScope 3/aScope 3 Slim to the aView and verify that 

a live video image appears on the screen.

  Always have an additional aScope at hand as a precaution.

1.

Functions

Warning: Read Instructions For Use before using the aScope

 3 or aScope

 3 Slim and aView.

1. & 2. Connect aScope 3 or aScope 3 Slim to the
aView by plugging the aScope 3 connector
- white with blue arrow - into the corresponding 
blue female connector on the side of aView.

3. Push the ON/OFF 
button on the aView to 
start up the system.

4. The real time image 
will appear on aView 
after only a few seconds 
and the aScope 3 system 
is ready to use

aScope 3

  Insertion cord diameter 5.0 mm
    Working channel 2.2 mm
      Bending angle 150º/130º

aScope 3 Slim

Insertion cord diameter 3.8 mm
Working channel 1.2 mm
Bending angle 130º/130º

Reviews: