5.19
Analog output option low value
Display:
REC_
Range:
Any display value
Default Value:
0
Seen only when analog retransmission option fitted. Refer to the separate “PM4 Panel Meter
Optional Output Addendum” booklet supplied when this option is fitted for wiring details and link
settings. Displays and sets the analog retransmission (4–20mA, 0–1V or 0–10V, link selectable)
output low value (4mA or 0V) in displayed engineering units. To set the analog output low value
go to the
REC_
function and use the
^
or
v
push buttons to set the required value then press
F
to accept this selection.
Example:
If it is required to retransmit 4mA when the display indicates
0
then select
0
in this
function using the
^
or
v
button.
5.20
Analog output option high value
Display:
REC~
Range:
Any display value
Default Value:
1000
Seen only when analog retransmission option fitted. Refer to the separate “PM4 Panel Meter
Optional Output Addendum” booklet supplied when this option is fitted for wiring details and link
settings. Displays and sets the analog retransmission (4–20mA, 0–1V or 0–10V, link selectable)
output high display value (20mA, 1V or 10V) in displayed engineering units. To set the analog
output high value go to the
REC~
function and use the
^
or
v
push buttons to set the required
value then press
F
to accept this selection.
Example:
If it is required to retransmit 20mA when the display indicates
50
then select
50
in
this function using the
^
or
v
button.
5.21
Display rounding
Display:
drnd
Range:
1
to
5000
Default Value:
1
Displays and sets the display rounding value. This value may be set to 1 - 5000 displayed units.
Display rounding is useful for reducing the instrument resolution without loss of accuracy in ap-
plications where it is undesirable to display to a fine tolerance. To set the display rounding value
go to the
drnd
function and use the
^
or
v
push buttons to set the required value then press
F
to accept this selection.
Example:
If set to
10
the display values will change in multiples of 10 only i.e. display moves
from
10
to
20
to
30
etc.
PM4SSIMAN-2.2-0