![ALTUS ClioAir Service Manual Download Page 14](http://html.mh-extra.com/html/altus/clioair/clioair_service-manual_2914371014.webp)
14
ALTUS
|
3731 Northridge Dr, NW
|
Suite 1
|
Walker, MI 49544
|
www.altus-inc.com |
AS-1584_Rev.I
Battery and Power System Removal and Replacement
DO NOT CONNECT THIS UNIT TO BATTERIES NOT SUPPLIED BY ALTUS
WITHOUT VERIFYING CONTROLLER SETUP AND OTHER OPTIONS WITH ALTUS.
WARNING!
If replacing the LiFe (Lithium Iron Phosphate) battery with a SLA battery or SLA to LiFe,
the following steps need to be completed. If the MPM software has been installed, this will help
with step #4.
1. All cables to and from controller must be connected, including the battery and its cables and connections.
Both the AC-In and AC-out connector ends include a locking feature to prevent them from pulling apart. To disconnect
pull back both “red” tabs.
Note: Older models may not have this feature (red tabs) so simply disconnect.
Controller must first be
OFF
(no lights on the Remote User Interface) and connected to AC power.
Perform this step only if using the MPM Software. If the MPM Software is not used, proceed to step 2.
Connect the USB cable into the technology component within the cart (or into a resident laptop with
the MPM software loaded).
2. Wait for the double yellow to appear on the interface.
3. When this happens, press and hold the mute button for 5 seconds. One “beep” will be made immediately.
Within 3-5 seconds after the initial “beep”, a second “beep” will be heard.
Keep holding down the mute button until a scrolling light sequence appears (release mute button at this time).
This will look like all of the green LED lights going lighting up in order and then back off in order. When green light
sequence is complete, the controller will show a Yellow LED and Green LED and then will shut itself off (all lights
will turn off) and then will “Reboot” itself. When controller reboots, the remote user interface will show the battery
level of the connected battery.
4. The “battery reading for confirmation” will only be available if MPM View is being utilized.
The cart is ready to be used.
Technology W
orkstation Solutions — Powered
air
No Lights
USB