D000348-00 Rev. B
11
Setup and Installation
An AlterG qualified technician will install the AlterG Anti-Gravity Treadmill after delivery. Please make sure
that you inspect the Anti-Gravity Treadmill upon delivery for any damage that may have occurred during
transportation. Take pictures and immediately report any damage to the shipping company and AlterG.
When you sign for the shipment of your Anti-Gravity Treadmill, you are taking responsibility for any damage
that may occur before installation.
Electrical Requirements
The recommended electrical outlet/power source for the AlterG Anti-Gravity Treadmill is a 20 ampere,
220VAC @ 50/60 Hz dedicated circuit with ground. The plug supplied with the AlterG Anti-Gravity Treadmill
is designated by the NEMA configuration system as 6-20P. The corresponding receptacle for the plug is a
NEMA 6-20R.
International systems are shipped with the NEMA 6-20P. An appropriate plug for the country and facility in
question should be wired as follows:
Black Conductor: Line
Black Conductor: Line
Green/Yellow Conductor: Ground
Customer
Circuit
Via 400
Rating
Cable (mm
2
)
Connector
Type
Region
20A
16A
4
LAPP USA # 321203
ANIXTER # 4A-1203EC-UL
US,
Canada
16A
12.8A
≥ 1.5
2.5
LAPP USA # 321203
ANIXTER # 4A-1203EC-UL
EU (DEU)
13A
10.4A
≥ 1.5
2.5
LAPP USA # 321203
ANIXTER # 4A-1203EC-UL
UK
32A
16A
2.5
LAPP USA # 321203
ANIXTER # 4A-1203EC-UL
No Plug Required on 20A Switch Spur
UK (ring)
32A
16A
≥ 4.0
IEC 60309 male plug
UK (ring)