Excell MCDSe series
Rel. 1
alsa
apparecchi medicali srl
11
Monopolar Function
50ohm
100ohm
150ohm
min max min max min max
Pure
Cut
150 1.68 1.44 1.48 1.28 1.35
Blend
1
1.64 1.68 1.44 1.46 1.37 1.39
Blend
2
1.40 1.46 1.04 1.06
Endo
1.35 1.49 1.19 1.28 1.11 1.21
Fulg
Forced
1.48 1.55 1.11 1.15 0.92 0.95
PinPoint
Contact 1.57 1.58 1.40 1.46 1.24 1.26
Monopolar Function
50ohm
100ohm
150ohm
Soft
1.50 1.53 1.32 1.36 1.21 1.26
Spray
1.41 1.46 1.04 1.06 0.88 0.89
Bipolar Function
10ohm
50ohm
100ohm
min max min max min max
Pure
Cut
2.58 2.7 1.51 1.56 1.13 1.18
Blend
2.34 2.42 1.44 1.48 1.07 1.10
Micro
2.39 2.43 1.44 1.47 1.07 1.10
Macro
2.39 2.43 1.44 1.46 1.07 1.10
CONTROL
OF
THE
H.F.
LEAKAGE
CURRENTS
Press the CAL button; when the displays blink, set the monopolar Cut section, as shown into the Figure:
MONOPOLAR
BIPOLAR
RAM
PURE
BLEND-1
FULG FORCED
PIN POINT CONTACT
PURE
BLEND
MICRO CV
MICRO HC
Press the CAL button in order to interrupt the displays blinking.
For each function of the equipment, it is necessary to adjust the limit of the R.F. leakage current. To do that,
connect the equipment under test to a system for the measurement of the high frequency leakage currents;
then, switch every single function and operate on the buttons of power increase/decrease, in order to bring
the registered value within the limits established by the Standard (lower than 150mA for monopolar – less
than 1% of maximum power for bipolar).
CALIBRATION OF THE OUTPUT CURRENT
Press the CAL button; when the displays blink, set the monopolar Cut and Coagulation section, as shown
into the Figure:
MONOPOLAR
BIPOLAR
RAM
PURE
BLEND-1
FULG FORCED
PIN POINT CONTACT
PURE
BLEND
MICRO CV
MICRO HC
Connect at the bipolar output a 200ohms non-inductive load with the current probe placed in series to the
load. Activate the bipolar Pure Cut mode. The setting values are reported in the following table. During the
switching on, operate on the buttons of power increase/decrease, in order to obtain the correspondent