background image

 

4

for the X face, which is to the left of the probe’s notch (notch = black triangle in the image). Suppose that the offset is 
presently correct. If so, then if the probe is laid on a nonmag-netic table with the X face pointing downward (i.e., with 
the X face in contact with the table), it will accurately display the vertical component of the Earth magnetic field at 
that location. In the northern hemisphere, this vertical component generally is downward (while the horizontal 
component points approximately to the north). In this configuration of the meter and probe, the X face is pointing 
downward and the meter is displaying the X value. Therefore the meter is presently reading the 

vertical

 component 

of the field. In the northern hemisphere, it should read a negative number (typically around -0.40 gauss). 
 
If the probe is then flipped 180° (as a pancake would be flipped) so that the X face is upward, then the meter should 
read the 

negative

 of the previous number. For example, if it had read -0.40 when facing downward, it should now 

read 0.40 (without a negative sign) when facing upward. Whenever there is no error in the zero offset, a single axis 
reading, when flipped 180° about that axis, should remain the same number except that the sign will reverse. 
 
Now suppose that the offset were 

not

 properly zeroed, at least in the X axis. Suppose that the X value is off by -0.15 

gauss, so in true zero field, X reads -0.15 gauss. This would mean in the above example that when the X face is 
upward, instead of reading 0.40 gauss, it would read 0.25 gauss. Similarly, when face-down it would read -0.55 
gauss (which is 0.15 gauss 

more negative

 than the true value of -0.40). Because 0.25 and -0.55 are not negatives of 

each other, we know that the offset is not set properly in this example. 
 
The OFFSET knob can be used to change the offset of each of X,Y,Z. However, these three offsets can only be 
modified while viewing X,Y, or Z on the “Measure” mode, and then pushing 

downward

 on the OFFSET knob so a 

triangle appears in the display over the words “Offset Adjust” on the label.  After turning the left knob to “Measure”, 
pressing “View” to select X, Y, or Z, and pressing down on the OFFSET knob so a triangle appears over the words 
“Offset Adjust”, then by rotating the OFFSET knob, you can change the offset on the axis (X, Y, or Z) that is being 
viewed. Turning clockwise will add; CCW will subtract. It is 0.01 gauss per click of the knob and there are 24 clicks 
per rotation. The effect of the offset is displayed immediately. For example, if X is stable and reading -0.15 gauss, 
and then the knob is turned 2 turns CW, the reading will become -0.15 + 0.48 = 0.33 gauss. You can press 

Reset

 

while in this mode to restore the factory offsets. Once these offsets are modified, they will be remembered until new 
offsets are entered.  
 
The triangle is removed by pressing OFFSET again, or by toggling the view back to “Magnitude” (magnitude does 
not technically have a its own offset; it is a computed variable taken from the measured values of X,Y,Z so its offset 
cannot be adjusted). The triangle is also removed if the left knob is turned to “Peak Hold” (or of course to OFF). 
 
To adjust for zero offset, view X and press down on OFFSET so the “Offset Adjust” triangle appears, as above. Lay 
the X face of the probe on the table (facing down) and note the reading. Then flip the probe 180° so the X face is up. 
Note this and then compute the average of the two numbers. This average is the offset error of X. For example, if the 
first number was -0.43 and the second was 0.43, then the average is zero and there is no offset error. If the numbers 
are instead 0.25 and -0.55 (as in a previous example), the average is -0.15 so that 0.15 would need to be added to 
the X offset to correct it. This amount is added by turning the OFFSET knob 15 clicks clockwise. If the two numbers 
are 1.50 and 1.70, the offset error is 1.60, so the knob must be turned CCW 6 complete rot 16 clicks. When 
offset is adjusted correctly, the two numbers will be the same amount but of opposite signs (subject to the 0.02 gauss 
jitter). Note and remember the final value when the X face is downward, which is the correct value of the vertical field 
component. Knowing this value will make the Y and Z offset corrections simpler to do. 
 
To correct Y for zero offset, press “View” to display Y and lay the Y face downward on the table. If the Y offset is 
correct, it will read the same number as the X did when X was facing downward (assuming the X offset was first 
corrected). Rotate the OFFSET knob until Y reads this number (the vertical field). You can confirm that the 
adjustment is correct by flipping Y so it is face-up; the number should reverse sign just as X did, and both numbers 
should be the same as the numbers for X. To correct Z, press “View” to display Z, lay the top surface of the probe 
(the Z face) against the table, so that the probe is vertical, and rotate OFFSET so that Z reads the same number as 
X did (when X was face-down). Now press downward on the OFFSET knob to remove the triangle, and the offset 
adjustment is done. (Accidentally turning the OFFSET knob will now not affect anything.) 

Reviews: