AlphaLab VGM Instructions Download Page 2

 

2

 

In theory, this number will not change when the probe is reoriented in a different direction. In practice, the reading 
will often change somewhat whenever the probe direction is reoriented. This problem occurs if the meter is either not 
properly “zeroed” or if the probe is near a small magnetized object which has a field that is not uniform across the 
thickness of the probe. (As shipped, the meter is properly zeroed to /- 0.10 gauss for each of X,Y,Z. Methods 
for altering and checking the meter zero are described later.) The field is not uniform if its source is a small part or if 
the field is strong only in a region that if a few millimeters wide. If you find a location on a magnetized surface where 
the field very strong and then move the probe away a few mm, the reading will not change much if the field is 
uniform. If the field magnitude instead changes by more than a factor of two over a distance of a few mm, then it is 
not uniform and it is more difficult to read consistently with a probe that is about 6 mm wide.  
 

Polarity of the probe:

 The X,Y,Z sensors read positive when the north pole of a magnet is pointed toward them. 

That is, if a magnet is placed on the right side with its north pole pointing to the left in the image above, then if the 
“View” button is pressed so that Z is displayed, the number shown will be positive (no negative sign will be present). 
If the magnet is then turned 180°, the number will become negative. Note that a south pole that is pointed toward the 
Z face will produce a negative value for Z. Also, if the Z face is pointed toward the Earth north pole, Y will read 

negative

-- the Earth 

north

 pole is actually the 

south

 pole of a magnet. 

 

Alarm:

 The alarm will sound (and a red light will appear from under the label) when the magnitude of the field 

exceeds a level that you set. The alarm will never sound or light unless you set a level, and it will not sound or light 
when muted (by setting it to 0.00). It will also become muted whenever the meter is turned off and back on. To set a 
level, Press 

and continue to press

 “Alarm Set”. The level will be displayed (initially 0.00) and can be changed by 

turning the OFFSET knob. Clockwise increases the number; counter-clockwise decreases it (one click CCW from 
0.00 is the highest alarm level: 790.00 gauss; more clicks CCW will decrease that level). At any time, the level can 
be directly reset to 0.00 (muted) by pushing 

down

 on the OFFSET knob while pressing “Alarm Set”. Once “Alarm 

Set” is released, the level will be remembered and can be modified later by pressing and holding “Alarm Set” again. 
There are about 100 different alarm levels that can be selected, from 1.00 to 790.00 (these values will be obvious). 
The alarm will produce sound and light only 

during

 the times that the magnitude exceeds the level you had set, and a 

delay of about 5 milliseconds is required for the alarm sound to begin. There is about a 5 millisecond lag (after the 

Reviews: