Alpha MED-A64HE1 Product Manual Download Page 9

7

Signal Acquisition with the MED64 system

 

The MED probe has 64 recording electrodes and 4 reference electrodes. The difference between potentials at the 
recording electrodes and the reference electrodes are measured with the MED64 system. Fig.1 shows a 
schematic diagram of the circuit between the recording and reference electrodes with the MED64 system.

Amplifier Input-Impedance Requirements for avoiding 
Signal Attenuation 

One of the planar microelectrode's unique characteristics is that it forms an “electric double layer capacitor” 
between the electrode and solution. This “electric double layer capacitor” results in a rise in the electrode 
impedance at low frequencies. (See Fig.2) This detail should be noted because acquired signals are attenuated if 
the electrode impedance becomes higher than the input impedance of the amplifier.

The platinum-black microelectrodes on the MED probe intrinsically have very low-impedances (at 1kHz, 10 k

 

for 50 

m electrodes, 40 k

 for 20 

m electrodes). The impedances are small even at low frequencies (as seen 

in Fig.2). Thus, the input-impedance of 100 M

 for the MED64 Head Amplifier is high enough for the MED64 

system. Low-frequency signals such as fEPSPs or even slower signals such as from smooth muscle (0.1 Hz) can 
be acquired without attenuation. 

If 1Hz signal is acquired with flat ITO electrodes, signals are attenuated because the electrode impedances are 
larger than 100 M

 (blue line in Fig. 2.). However, when 1Hz signals are acquired with the platinum-black 

electrode on the MED probe, signals are acquired without attenuation because the electrode impedances are 
smaller than 100 M

 (green and red lines in Fig.2).

Fig.1. 

Electrodes embedded on the MED probe (left) and the mechanism of recording of field potentials with the MED64

Reference electrodes

system (right).

(100 M

)

MED64 electrode (Pt. black. 50 

m)

 

ITO electrode (50 

m) 

MED64 electrode (Pt. black. 20 

m)

Fig.2.

 Electrode impedance and input impedance of the MED64 Head Amplifier.

Input impedance of MED64 Head Amplifier 

             Attenuation

Summary of Contents for MED-A64HE1

Page 1: ...Product Manual MED64 Head Amplifier 64 CHANNEL HEAD AMPLIFIER P N MED A64HE1 A low noise multi electrode array system for in vitro electrophysiology...

Page 2: ...mages resulting from the use of information contained in this document or from the use of programs and source code that may accompany it In no event shall the publisher and the author be liable for an...

Page 3: ...e MED64 Main Amplifier to other Components 6 Signal Acquisition with the MED64 system 7 Amplifier Input Impedance Requirements for avoiding Signal Attenuation 7 Stimulation Using Planar Microelectrode...

Page 4: ...ower cords with cuts scratches or poor connectors as this may result in fire or electric shock Excessive bending pulling or slicing of the cord should also be avoided Do not pull on the cord when you...

Page 5: ...symbols used on equipment and documentation Environmental Conditions Indoor use Altitude up to 2000 m Temperature 5 40 C Maximum relative humidity 80 for temperatures up to 31 C decreasing linearly to...

Page 6: ...livered via any of the 64 electrodes using the built in stimulator in the MED64 Head Amplifier Components and their functions Front panel 1 POWER Turns amplifier ON and OFF 2 STIMULUS CURRENT Selectio...

Page 7: ...2 STIMULUS OUTPUT on the 64 CHANNEL MAIN AMPLIFIER MED A64MD1 5 CONTROL INPUT Input for control of stimulus channels Connects to the CONTROL OUTPUT on the 64 CHANNEL MAIN AMPLIFIER MED A64MD1 6 SIGNAL...

Page 8: ...idified incubator It is connected with a cable as long as 2m without noise or signal attenuation The MED64 system does NOT usually require a Faraday cage or vibration isolation table necessary for con...

Page 9: ...have very low impedances at 1kHz 10 k for 50 m electrodes 40 k for 20 m electrodes The impedances are small even at low frequencies as seen in Fig 2 Thus the input impedance of 100 M for the MED64 Hea...

Page 10: ...field potentials Vf change according to the output current Is as seen on the right in the Fig 4 Stimulus Artifacts and Biphasic Stimulation During stimulation signals stimulus pulses are recorded at...

Page 11: ...ositive components as seen in Fig 5 When biphasic stimulation is applied the electrode is charged and discharged quickly However when mono phasic stimulation is applied the electrodes need longer time...

Page 12: ...s 50 000 pF for 50 x 50 m electrodes 22 000 pF for 20 x 20 m electrodes allowing stimulation with current amplitudes up to 200 A for 0 2 msec duration for each phase is 0 1 msec DO NOT use stimulus cu...

Page 13: ...n maladjustment of user controls improper maintenance modifications or service by anyone other than AMS or damage attributable to acts of God Specifications Amplifier Stimulator General Power supply u...

Page 14: ...ntific Inc Saito Bio Incubator 209 7 7 15 Saito asagi Ibaraki Osaka 567 0085 Japan Phone 81 72 648 7973 FAX 81 72 648 7974 http www med64 com support med64 com Manufactured by Alpha MED Scientific Inc...

Reviews: